» Articles » PMID: 16760231

Remote Detection of Incessant Slow VT with an ICD Capable of Home Monitoring

Overview
Journal Europace
Date 2006 Jun 9
PMID 16760231
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

In this case report, we present a 57-year-old patient who suffered from incessant slow ventricular tachycardia because of a change in therapy 2 weeks after the implantation of an implantable cardioverter defibrillator (ICD) with Home Monitoring (HM). HM is a fully automatic and wireless system capable of transmitting device data including episode counters. The patient had mild symptoms with only short episodes of dizziness and would not have contacted the primary care centre. Due to HM, the arrhythmia was detected before the next routine follow-up, and the patient was seen the next day. HM made it possible to adjust the dosage of the newly established beta-blocker treatment in an outpatient setting and avoid hospitalization. HM yields the potential for remote detection of arrhythmias in ICD recipients, allowing alterations in device programming or medication.

Citing Articles

[Home monitoring of an incessant VT in an ICD patient].

Reinsch N, Woydowski D, Schon H, Buddensiek M, Weissenberger W, Erbel R Herz. 2010; 36(1):28-32.

PMID: 21181096 DOI: 10.1007/s00059-010-3411-6.


Advances in remote monitoring of implantable pacemakers, cardioverter defibrillators and cardiac resynchronization therapy systems.

Jung W, Rillig A, Birkemeyer R, Miljak T, Meyerfeldt U J Interv Card Electrophysiol. 2008; 23(1):73-85.

PMID: 18821006 DOI: 10.1007/s10840-008-9311-5.