» Articles » PMID: 16646048

Automated Fluorescent in Situ Hybridization (FISH) Analysis of T(9;22)(q34;q11) in Interphase Nuclei

Overview
Journal Cytometry A
Specialties Cell Biology
Radiology
Date 2006 Apr 29
PMID 16646048
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Background: For chronic myeloid leukemia, the FISH detection of t(9;22)(q34;q11) in interphase nuclei of peripheral leukocytes is an alternative method to bone marrow karyotyping for monitoring treatment. With automation, several drawbacks of manual analysis may be circumvented. In this article, the capabilities of a commercially available automated image acquisition and analysis system were determined by detecting t(9;22)(q34;q11) in interphase nuclei of peripheral leukocytes.

Methods: Three peripheral blood samples of normal adults, 21 samples of CML patients, and one sample of a t(9;22)(q34;q11) positive cell-line were used.

Results: Single nuclei with correctly detected signals amounted to 99.6% of nuclei analyzed after exclusion of overlapping nuclei and nuclei with incorrect signal detection. A cut-off value of 0.84 mum was defined to discriminate between translocation positive and negative nuclei based on the shortest distance between signals. Using this value, the false positive rate of the automated analysis for negative samples was 7.0%, whereas that of the manual analysis was 5.8%. Automated and manual results showed strong correlation (R(2) = 0.985), the mean difference of results was only 3.7%.

Conclusions: A reliable and objective automated analysis of large numbers of cells is possible, avoiding interobserver variability and producing statistically more accurate results than manual evaluation.

Citing Articles

Automated 3D scoring of fluorescence in situ hybridization (FISH) using a confocal whole slide imaging scanner.

Frankenstein Z, Uraoka N, Aypar U, Aryeequaye R, Rao M, Hameed M Appl Microsc. 2021; 51(1):4.

PMID: 33835321 PMC: 8035347. DOI: 10.1186/s42649-021-00053-y.


Feature selection for the automated detection of metaphase chromosomes: performance comparison using a receiver operating characteristic method.

Qiu Y, Song J, Lu X, Li Y, Zheng B, Li S Anal Cell Pathol (Amst). 2015; 2014:565392.

PMID: 25763334 PMC: 4334018. DOI: 10.1155/2014/565392.


Investigation of CPD and HMDS sample preparation techniques for cervical cells in developing computer-aided screening system based on FE-SEM/EDX.

Jusman Y, Ng S, Abu Osman N ScientificWorldJournal. 2015; 2014:289817.

PMID: 25610902 PMC: 4291196. DOI: 10.1155/2014/289817.


Intelligent screening systems for cervical cancer.

Jusman Y, Ng S, Abu Osman N ScientificWorldJournal. 2014; 2014:810368.

PMID: 24955419 PMC: 4037632. DOI: 10.1155/2014/810368.


Fluorescence in situ hybridization (FISH): an increasingly demanded tool for biomarker research and personalized medicine.

Hu L, Ru K, Zhang L, Huang Y, Zhu X, Liu H Biomark Res. 2014; 2(1):3.

PMID: 24499728 PMC: 3917523. DOI: 10.1186/2050-7771-2-3.