Stability of the EasyOne Ultrasonic Spirometer for Use in General Practice

Overview
Journal Respirology
Specialty Pulmonary Medicine
Date 2006 Apr 26
PMID 16635089
Citations 25
Authors
Affiliations
Soon will be listed here.
Abstract

Objective And Background: Spirometry is recommended for the diagnosis and management of chronic respiratory diseases in the community. Spirometer accuracy is critical, but few general practitioners meet the American Thoracic Society and European Respiratory Society (ATS/ERS) recommendation for daily calibration. The aim of this study was to assess the accuracy and stability of a portable ultrasonic spirometer (EasyOne) that the manufacturer claims does not require regular calibration.

Methods: Six EasyOne spirometers were used in a practice-based spirometry study. Inspiratory and expiratory accuracy was checked periodically using a certified 3-L syringe. Paired calibration checks were performed using a dedicated mouthpiece assembly (spirette) and randomly selected spirettes.

Results: The six spirometers were used for up to 26 weeks (mean 23.9 weeks) and a total of 1041 spirometry tests and 75 syringe calibrations were performed. All inspiratory and expiratory calibration checks using a dedicated or randomly selected spirette met the ATS/ERS accuracy criterion (3.00+/-0.105 L). The mean (range) expiratory volume deviation from target volume (3.00 L) was 0.011 L (-0.06-0.09 L) using a dedicated spirette and 0.046 L (-0.09-0.10 L) using randomly selected spirettes. The deviation from target was not affected by the mean flow generated during the calibration procedure. There was no change in calibration during the study.

Conclusions: This study supports the manufacturer's claim that the EasyOne spirometer maintains its calibration during routine clinical use in general practice and does not require daily calibration as specified in international spirometry guidelines.

Citing Articles

Exhaled Breath Analysis: From Laboratory Test to Wearable Sensing.

Heng W, Yin S, Chen Y, Gao W IEEE Rev Biomed Eng. 2024; 18:50-73.

PMID: 39412981 PMC: 11875904. DOI: 10.1109/RBME.2024.3481360.


Validity of a portable spirometer in the communities of China.

Xiao S, Wu F, Wang Z, Chen J, Yang H, Zheng Y BMC Pulm Med. 2022; 22(1):80.

PMID: 35248001 PMC: 8898436. DOI: 10.1186/s12890-022-01872-9.


Decreased Pulmonary Function Over 5 Years in US Firefighters.

Mathias K, Graham E, Stewart D, Smith D J Occup Environ Med. 2020; 62(10):816-819.

PMID: 32590426 PMC: 7537734. DOI: 10.1097/JOM.0000000000001944.


Increased pulmonary artery diameter is associated with reduced FEV in former World Trade Center workers.

de la Hoz R, Jeon Y, Reeves A, San Jose Estepar R, Liu X, Doucette J Clin Respir J. 2019; 13(10):614-623.

PMID: 31347281 PMC: 6783324. DOI: 10.1111/crj.13067.


Joint Indian Chest Society-National College of Chest Physicians (India) guidelines for spirometry.

Aggarwal A, Agarwal R, Dhooria S, Prasad K, Sehgal I, Muthu V Lung India. 2019; 36(Supplement):S1-S35.

PMID: 31006703 PMC: 6489506. DOI: 10.4103/lungindia.lungindia_300_18.