» Articles » PMID: 16565006

Ultrasonic-generated Fluid Velocity with Sovereign WhiteStar Micropulse and Continuous Phacoemulsification

Overview
Specialty Ophthalmology
Date 2006 Mar 28
PMID 16565006
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate and compare ultrasonic turbulence created by conventional and micropulse ultrasound technology.

Setting: Sonora Medical Systems, Longmont, Colorado, USA.

Methods: A high-resolution digital ultrasound probe imaged the zone around a phacoemulsification tip. Doppler analysis allowed determination of flow. The fluid velocity was measured at 4 levels of ultrasound power at a constant flow, comparing the ultrasonic conditions of continuous energy to WhiteStar micropulses.

Results: In addition to the normal baseline irrigation and aspiration, fluid movement was detected directly below the phaco tip, produced by a nonlinear effect known as acoustic streaming. Acoustic streaming increased with increased phacoemulsification power for both conditions. At each of the 4 levels of power, fluid velocity away from the tip was less with micropulse technology than with continuous phacoemulsification.

Conclusions: The demonstrated decrease in acoustic streaming flow away from the phaco tip with Sovereign WhiteStar micropulse technology compared to conventional ultrasound provides an objective explanation for clinical observations of increased stability of nuclear fragments at the tip and less turbulence in the anterior chamber during phacoemulsification. This methodology can be used to examine and compare fluid flow and turbulence under a variety of clinically relevant conditions.

Citing Articles

Simulation of fluid dynamics and turbulence during phacoemulsification using the new propeller turbo tip.

Wang X, Wang X BMJ Open Ophthalmol. 2023; 8(1).

PMID: 37775137 PMC: 10546181. DOI: 10.1136/bmjophth-2023-001391.


Effects of Cataract Surgery on Endothelium in Transplanted Corneal Grafts: Comparison of Extracapsular Cataract Extraction and Phacoemulsification for Complicated Cataract after Penetrating Keratoplasty.

Zhou H, Xie L Chin Med J (Engl). 2016; 129(17):2096-101.

PMID: 27569238 PMC: 5009595. DOI: 10.4103/0366-6999.189050.


Visualization of irrigation fluid flow and calculation of its velocity distribution in the anterior chamber by particle image velocimetry.

Kaji Y, Yamashita M, Sakakibara J, Oshika T Graefes Arch Clin Exp Ophthalmol. 2012; 250(7):1023-7.

PMID: 22350059 DOI: 10.1007/s00417-012-1953-4.