» Articles » PMID: 16547332

Retention and Removal of a New Viscous Dispersive Ophthalmic Viscosurgical Device During Cataract Surgery in Animal Eyes

Overview
Journal Br J Ophthalmol
Specialty Ophthalmology
Date 2006 Mar 21
PMID 16547332
Citations 18
Authors
Affiliations
Soon will be listed here.
Abstract

Aims: To assess the retention and removal properties of a new viscous dispersive ophthalmic viscosurgical device (OVD), DisCoVisc, in comparison with those of cohesive (Provisc), dispersive (Viscoat), and viscoadaptive (Healon5) OVDs.

Methods: In 20 porcine eyes, cataract surgery was simulated using one of the four OVDs which were stained with fluorescein for better visualisation. Three parameters were measured. Firstly, the presence/absence of OVDs in the chamber at the completion of phacoemulsification was recorded. Secondly, the time until the OVDs were completely removed from the anterior chamber using the phaco needle was measured. Thirdly, after intraocular lens (IOL) implantation, the time needed to completely remove the OVDs from the chamber with irrigation/aspiration tip was recorded.

Results: At the completion of phacoemulsification, the OVDs retained in 0% (0/5) for Provisc, 80% (4/5) for Healon5, 100% (5/5) for DisCoVisc, and 100% (5/5) for Viscoat. The retention of OVDs during phacoemulsification was greatest with Viscoat followed by, in descending order, DisCoVisc, Healon5, and Provisc. The removal of OVDs after IOL implantation took longest with Viscoat followed by Healon5, DisCoVisc, and Provisc.

Conclusion: The viscous dispersive DisCoVisc showed excellent retention during phacoemulsification, while its removal after IOL implantation was very easy. When compared with the viscoadaptive Healon5, DisCoVisc was retained better in the chamber and was easier to remove. These features of DisCoVisc should be highly advantageous when considering covering the entire cataract surgery procedure with a single OVD.

Citing Articles

Physical Characteristics, Clinical Application, and Side Effects of Viscoelastics in Ophthalmology.

Mohammadpour M, Asadigandomani H, Aminizade M, Raeisi S J Curr Ophthalmol. 2024; 35(4):313-319.

PMID: 39281404 PMC: 11392293. DOI: 10.4103/joco.joco_178_23.


Comparison of the Residual Amount of an Ophthalmic Viscosurgical Device Among Different Types of Intraocular Lens Implants in vitro.

Oki T, Horiguchi H, Terauchi R, Gunji H, Nakano T Clin Ophthalmol. 2024; 18:1083-1091.

PMID: 38659426 PMC: 11041993. DOI: 10.2147/OPTH.S458348.


Temperature-controlled porcine eye holder for observing intraocular temperature during cataract surgery.

Minami K, Yaguchi S, Bissen-Miyajima H Sci Rep. 2023; 13(1):4331.

PMID: 36922583 PMC: 10017689. DOI: 10.1038/s41598-023-31070-4.


Thickness of the Protective Layers of Different Ophthalmic Viscosurgical Devices During Lens Surgery in a Porcine Model.

Wust M, Matten P, Nenning M, Findl O Transl Vis Sci Technol. 2022; 11(2):28.

PMID: 35175318 PMC: 8857612. DOI: 10.1167/tvst.11.2.28.


Ophthalmic Viscosurgical Devices (OVDs) in Challenging Cases: a Review.

Borkenstein A, Borkenstein E, Malyugin B Ophthalmol Ther. 2021; 10(4):831-843.

PMID: 34617249 PMC: 8589875. DOI: 10.1007/s40123-021-00403-9.


References
1.
Assia E, Apple D, Lim E, Morgan R, Tsai J . Removal of viscoelastic materials after experimental cataract surgery in vitro. J Cataract Refract Surg. 1992; 18(1):3-6. DOI: 10.1016/s0886-3350(13)80376-8. View

2.
Poyer J, Chan K, Arshinoff S . New method to measure the retention of viscoelastic agents on a rabbit corneal endothelial cell line after irrigation and aspiration. J Cataract Refract Surg. 1998; 24(1):84-90. DOI: 10.1016/s0886-3350(98)80079-5. View

3.
McDermott M, Hazlett L, Barrett R, Lambert R . Viscoelastic adherence to corneal endothelium following phacoemulsification. J Cataract Refract Surg. 1998; 24(5):678-83. DOI: 10.1016/s0886-3350(98)80265-4. View

4.
Arshinoff S . Dispersive-cohesive viscoelastic soft shell technique. J Cataract Refract Surg. 1999; 25(2):167-73. DOI: 10.1016/s0886-3350(99)80121-7. View

5.
Auffarth G, Holzer M, Visessook N, Apple D, Volcker H . Removal times for a dispersive and a cohesive ophthalmic viscosurgical device correlated with intraocular lens material. J Cataract Refract Surg. 2004; 30(11):2410-4. DOI: 10.1016/j.jcrs.2004.03.039. View