[MRI Safety Aspects and Artifacts of Atrial Septal Defect and Patent Foramen Ovale Occluders at 1.5 Tesla: a Phantom Study]

Overview
Journal Rofo
Publisher Thieme
Date 2006 Mar 2
PMID 16508833
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To investigate the magnetic resonance safety and magnetic resonance artifact intensities of patent foramen ovale and atrial septal occluders in a phantom study.

Materials And Methods: Seven different commercial occluder models (Amplatzer ASD, Angel Wing ASD, Helex Septal Occluder, Cardia PFO-Star Generation I & III, Sideris Buttoned Device, Starflex ASD) were tested in a 1.5 Tesla whole-body MR scanner. Following a deflection test, the non-magnetic devices were imaged using different MRI pulse sequences including conventional spin echo, spoiled gradient echo as well as fast refocused steady-state pulse sequences (trueFISP) commonly used for cardiac MRI. Additionally, device temperatures were measured in the MR scanner during two imaging protocols (trueFISP, HASTE) with an MR-compatible fiber-optical thermometer.

Results: One ferromagnetic device (Sideris Buttoned Device) did not pass the deflection test and was excluded from further experiments. For typical cardiac imaging protocols, the remaining non-ferromagnetic devices showed only minor image artifacts with artifact sizes not exceeding 2 mm next to a strut. Heating was not observed in the devices.

Conclusion: Non-ferromagnetic occluder models used in this study can safely be imaged with typical cardiac MRI protocols at 1.5 Tesla. The device artifact sizes are small enough to allow anatomical and functional MRI even in the immediate vicinity of the occluders.

Citing Articles

Responsivity of the Striatal Dopamine System to Methylphenidate-A Within-Subject I-123-β-CIT-SPECT Study in Male Children and Adolescents With Attention-Deficit/Hyperactivity Disorder.

Aster H, Romanos M, Walitza S, Gerlach M, Muhlberger A, Rizzo A Front Psychiatry. 2022; 13:804730.

PMID: 35492708 PMC: 9046584. DOI: 10.3389/fpsyt.2022.804730.


Contrast-enhanced CMR in patients after percutaneous closure of the left atrial appendage: a pilot study.

Mohrs O, Wunderlich N, Petersen S, Pottmeyer A, Kauczor H J Cardiovasc Magn Reson. 2011; 13:33.

PMID: 21726450 PMC: 3146401. DOI: 10.1186/1532-429X-13-33.


ACCF/ACR/AHA/NASCI/SCMR 2010 expert consensus document on cardiovascular magnetic resonance: a report of the American College of Cardiology Foundation Task Force on Expert Consensus Documents.

Hundley W, Bluemke D, Finn J, Flamm S, Fogel M, Friedrich M J Am Coll Cardiol. 2010; 55(23):2614-62.

PMID: 20513610 PMC: 3042771. DOI: 10.1016/j.jacc.2009.11.011.


ACCF/ACR/AHA/NASCI/SCMR 2010 expert consensus document on cardiovascular magnetic resonance: a report of the American College of Cardiology Foundation Task Force on Expert Consensus Documents.

Hundley W, Bluemke D, Finn J, Flamm S, Fogel M, Friedrich M Circulation. 2010; 121(22):2462-508.

PMID: 20479157 PMC: 3034132. DOI: 10.1161/CIR.0b013e3181d44a8f.


Atrial septal defects type II: noninvasive evaluation of patients before implantation of an Amplatzer Septal Occluder and on follow-up by magnetic resonance imaging compared with TEE and invasive measurement.

Weber C, Weber M, Ekinci O, Neumann T, Deetjen A, Rolf A Eur Radiol. 2008; 18(11):2406-13.

PMID: 18568349 DOI: 10.1007/s00330-008-1033-7.