» Articles » PMID: 16441172

Parallel Measurements of Drug Actions on Erythrocytes by Dielectrophoresis, Using a Three-dimensional Electrode Design

Overview
Specialty Biotechnology
Date 2006 Jan 31
PMID 16441172
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

A type of well-based assay that uses a laminated three-dimensional electrode design to characterise the effects of different drugs on red blood cells using dielectrophoresis is presented. The capability of the system to determine the effects of chemical agents on the electrophysiology of red blood cells is demonstrated using saponin and valinomycin as two examples of drugs that can penetrate the cell membrane and therefore change the dielectric properties of the cell. Light intensity changes are measured in the well over a period of time at various frequencies and the dielectric properties of the cells determined using an ellipsoidal multi-shell model. It is shown that the laminated electrode permits a high degree of automation and thus a high number of parallel experiments, which reduces both the time and effort needed to examine differences between populations of red blood cells. The technique is directly compatible with the industry-standard 1536 well-plate analysis technique.

Citing Articles

Evaluating refrigeration and antibiotic treatment for maintaining urine electrophysiology.

Hoque R, Labeed F, Cirovic S, Hughes M PLoS One. 2025; 20(2):e0319089.

PMID: 39999201 PMC: 11856264. DOI: 10.1371/journal.pone.0319089.


Electrical Impedance Spectroscopy as a Tool to Detect the Epithelial to Mesenchymal Transition in Prostate Cancer Cells.

Simpkins L, Henriquez L, Tran M, Adams T Biosensors (Basel). 2024; 14(10).

PMID: 39451716 PMC: 11506005. DOI: 10.3390/bios14100503.


Label-free, non-contact determination of resting membrane potential using dielectrophoresis.

Hughes M, Clarke K, Hoque R, Griffiths O, Kruchek E, Johnson M Sci Rep. 2024; 14(1):18477.

PMID: 39122771 PMC: 11316104. DOI: 10.1038/s41598-024-69000-7.


Ten-Second Electrophysiology: Evaluation of the 3DEP Platform for high-speed, high-accuracy cell analysis.

Hoettges K, Henslee E, Torcal Serrano R, Jabr R, Abdallat R, Beale A Sci Rep. 2019; 9(1):19153.

PMID: 31844107 PMC: 6915758. DOI: 10.1038/s41598-019-55579-9.


Isolation and concentration of bacteria from blood using microfluidic membraneless dialysis and dielectrophoresis.

DAmico L, Ajami N, Adachi J, Gascoyne P, Petrosino J Lab Chip. 2017; 17(7):1340-1348.

PMID: 28276545 PMC: 5894507. DOI: 10.1039/c6lc01277a.