» Articles » PMID: 16319085

Rapid Online Quantification of Left Ventricular Volume from Real-time Three-dimensional Echocardiographic Data

Overview
Journal Eur Heart J
Date 2005 Dec 2
PMID 16319085
Citations 76
Authors
Affiliations
Soon will be listed here.
Abstract

Aims: Determination of left ventricular (LV) volumes and ejection fraction (EF) from two-dimensional echocardiographic (2DE) images is subjective, time-consuming, and relatively inaccurate because of foreshortened views and the use of geometric assumptions. Our aims were (1) to validate a new method for rapid, online measurement of LV volumes from real-time three-dimensional echocardiographic (RT3DE) data using cardiac magnetic resonance (CMR) as the reference and (2) to compare its accuracy and reproducibility with standard 2DE measurements.

Methods And Results: CMR, 2DE, and RT3DE datasets were obtained in 50 patients. End-systolic and end-diastolic volumes (ESV and EDV) were calculated from the 2DE images using biplane method of disks. ES and ED RT3DE datasets were analysed using prototype software designed to automatically detect the endocardial surface using a deformable shell model and calculate ESV and EDV from voxel counts. 2DE and RT3DE-derived volumes were compared with CMR (linear regression, Bland-Altman analysis). In most patients, analysis of RT3DE data required <2 min per patient. RT3DE measurements correlated highly with CMR (r: 0.96, 0.97, and 0.93 for EDV, ESV, and EF, respectively) with small biases (-14 mL, -6.5 mL, -1%) and narrow limits of agreement (SD: 17 mL, 16 mL, 6.4%). 2DE measurements correlated less well with CMR (r: 0.89, 0.92, 0.86) with greater biases (-23 mL, -15 mL, 1%) and wider limits of agreement (SD: 29 mL, 24 mL, 9.5%). RT3DE resulted in lower intra-observer (EDV: 7.9 vs. 23%; ESV: 7.6 vs. 26%) and inter-observer variability (EDV: 11 vs. 26%; ESV: 13 vs. 31%).

Conclusion: Semi-automated detection of the LV endocardial surface from RT3DE data is suitable for clinical use because it allows rapid, accurate, and reproducible measurements of LV volumes, superior to conventional 2DE methods.

Citing Articles

Three-Dimensional Modelling of Indexed Papillary Muscle Displacement in Patients Requiring Mitral Valve Surgery Using Four-Dimensional Echocardiography Variables.

Ong Z, Ler A, Shen L, Kofidis T, Ti L, Sazzad F J Clin Med. 2025; 13(24.

PMID: 39768426 PMC: 11677786. DOI: 10.3390/jcm13247503.


Evidence of Left Ventricular Cardiac Remodeling After 6 Weeks of Sprint Interval Training.

Eriksson L, Hedman K, Astrom-Aneq M, Nylander E, Bouma K, Mandic M Scand J Med Sci Sports. 2024; 34(12):e70007.

PMID: 39707626 PMC: 11662151. DOI: 10.1111/sms.70007.


Validation of Volume Calibration by Echocardiography for Invasive Ventricular Pressure Volume Studies in Transcatheter Valve Interventions.

van den Dorpel M, van den Enden A, Verhemel S, Adrichem R, Ren C, Kardys I Struct Heart. 2024; 8(4):100307.

PMID: 39100581 PMC: 11294891. DOI: 10.1016/j.shj.2024.100307.


Anthracyclines-Induced Cardiac Dysfunction: What Every Clinician Should Know.

Ferrera A, Fiorentini V, Reale S, Solfanelli G, Tini G, Barbato E Rev Cardiovasc Med. 2024; 24(5):148.

PMID: 39076747 PMC: 11273047. DOI: 10.31083/j.rcm2405148.


HER2/neu 655 polymorphism, trastuzumab-induced cardiotoxicity, and survival in HER2-positive breast cancer patients.

Blancas I, Linares-Rodriguez M, Martin-Bravo C, Gomez-Pena C, Rodriguez-Serrano F Clin Transl Oncol. 2024; 26(10):2531-2540.

PMID: 38771533 PMC: 11410839. DOI: 10.1007/s12094-024-03512-6.