» Articles » PMID: 16244841

Reliability of Near-infrared Spectroscopy in People with Dark Skin Pigmentation

Overview
Publisher Springer
Date 2005 Oct 26
PMID 16244841
Citations 34
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: Near-infrared spectroscopy (NIRS) is a promising non-invasive technique for the continuous monitoring of tissue oxygen delivery. NIRS detects light absorbance of haemoglobin chromophores to determine tissue oxygen saturation (StO2). As skin colour is also determined by the presence of chromophores, it is plausible that NIRS signal quality may be affected by dark skin pigmentation.

Methods: Tissue saturation in the anterior compartment of the lower leg during isometric contraction was measured using NIRS in 17 volunteers with dark skin pigmentation. Measurements were continued until StO2 was zero percent or until the signal disappeared.

Results: The NIRS device failed to register tissue saturation values at some point in nine of seventeen volunteers. This occurred more often in individuals with darker skin.

Conclusions: In patients with a dark pigmented skin, NIRS StO2 measurements should be interpreted with caution, as melanin clearly interferes with the quality of the reflected NIRS signal.

Citing Articles

Challenging the skin pigmentation bias in tissue oximetry via time-domain near-infrared spectroscopy.

Lacerenza M, Amendola C, Bargigia I, Bossi A, Buttafava M, Calcaterra V Biomed Opt Express. 2025; 16(2):690-708.

PMID: 39958842 PMC: 11828448. DOI: 10.1364/BOE.541239.


Estimation of Pressure Pain in the Lower Limbs Using Electrodermal Activity, Tissue Oxygen Saturation, and Heart Rate Variability.

Kim Y, Pyo S, Lee S, Park C, Song S Sensors (Basel). 2025; 25(3).

PMID: 39943319 PMC: 11821113. DOI: 10.3390/s25030680.


Development and characterization of silicone-based tissue phantoms for pulse oximeter performance testing.

Bhusal A, Farahmand M, Hasan M, Vasudevan S, Vogt W, Ibarra B J Biomed Opt. 2025; 29(Suppl 3):S33314.

PMID: 39776836 PMC: 11706025. DOI: 10.1117/1.JBO.29.S3.S33314.


Near-infrared diffuse optical characterization of human thyroid using ultrasound-guided hybrid time-domain and diffuse correlation spectroscopies.

Fernandez Esteberena P, Cortese L, Zanoletti M, Lo Presti G, Aranda Velazquez G, Ruiz Janer S Biomed Opt Express. 2024; 15(12):7035-7055.

PMID: 39679411 PMC: 11640565. DOI: 10.1364/BOE.538141.


Exploring the impact and influence of melanin on frequency-domain near-infrared spectroscopy measurements.

Roy S, Wu J, Cao J, Disu J, Bharadwaj S, Meinert-Spyker E J Biomed Opt. 2024; 29(Suppl 3):S33310.

PMID: 39323492 PMC: 11423252. DOI: 10.1117/1.JBO.29.S3.S33310.


References
1.
Breit G, GROSS J, Watenpaugh D, Chance B, Hargens A . Near-infrared spectroscopy for monitoring of tissue oxygenation of exercising skeletal muscle in a chronic compartment syndrome model. J Bone Joint Surg Am. 1997; 79(6):838-43. DOI: 10.2106/00004623-199706000-00006. View

2.
Arbabi S, Brundage S, Gentilello L . Near-infrared spectroscopy: a potential method for continuous, transcutaneous monitoring for compartmental syndrome in critically injured patients. J Trauma. 1999; 47(5):829-33. DOI: 10.1097/00005373-199911000-00002. View

3.
Boushel R, Piantadosi C . Near-infrared spectroscopy for monitoring muscle oxygenation. Acta Physiol Scand. 2000; 168(4):615-22. DOI: 10.1046/j.1365-201x.2000.00713.x. View

4.
McCormick P, Stewart M, Goetting M, Balakrishnan G . Regional cerebrovascular oxygen saturation measured by optical spectroscopy in humans. Stroke. 1991; 22(5):596-602. DOI: 10.1161/01.str.22.5.596. View

5.
McCormick P, Stewart M, Goetting M, Dujovny M, Lewis G, Ausman J . Noninvasive cerebral optical spectroscopy for monitoring cerebral oxygen delivery and hemodynamics. Crit Care Med. 1991; 19(1):89-97. DOI: 10.1097/00003246-199101000-00020. View