Abstract

The goal of retinal prosthetic devices is to generate meaningful visual information in patients that have lost outer retinal function. To accomplish this, these devices should generate patterns of ganglion cell activity that closely resemble the spatial and temporal components of those patterns that are normally elicited by light. Here, we developed a stimulus paradigm that generates precise temporal patterns of activity in retinal ganglion cells, including those patterns normally generated by light. Electrical stimulus pulses (> or =1-ms duration) elicited activity in neurons distal to the ganglion cells; this resulted in ganglion cell spiking that could last as long as 100 ms. However, short pulses, <0.15 ms, elicited only a single spike within 0.7 ms of the leading edge of the pulse. Trains of these short pulses elicited one spike per pulse at frequencies < or =250 Hz. Patterns of short electrical pulses (derived from normal light elicited spike patterns) were delivered to ganglion cells and generated spike patterns that replicated the normal light patterns. Finally, we found that one spike per pulse was elicited over almost a 2.5:1 range of stimulus amplitudes. Thus a common stimulus amplitude could accommodate a 2.5:1 range of activation thresholds, e.g., caused by differences arising from cell biophysical properties or from variations in electrode-to-cell distance arising when a multielectrode array is placed on the retina. This stimulus paradigm can generate the temporal resolution required for a prosthetic device.

Citing Articles

Spatially Localized Visual Perception Estimation by Means of Prosthetic Vision Simulation.

Villarreal D, Krautschneider W J Imaging. 2024; 10(11).

PMID: 39590758 PMC: 11595353. DOI: 10.3390/jimaging10110294.


Comparison of modulation efficiency between normal and degenerated primate retina.

Yoo Y, Cha S, Goo Y Front Cell Dev Biol. 2024; 12:1419007.

PMID: 39144253 PMC: 11322106. DOI: 10.3389/fcell.2024.1419007.


Membrane depolarization mediates both the inhibition of neural activity and cell-type-differences in response to high-frequency stimulation.

Lee J, Werginz P, Kameneva T, Im M, Fried S Commun Biol. 2024; 7(1):734.

PMID: 38890481 PMC: 11189419. DOI: 10.1038/s42003-024-06359-3.


The Variation of Electrical Pulse Duration Elicits Reliable Network-Mediated Responses of Retinal Ganglion Cells in Normal, Not in Degenerate Primate Retinas.

Cha S, Ahn J, Kim S, Choi K, Yoo Y, Eom H Bioengineering (Basel). 2023; 10(10).

PMID: 37892865 PMC: 10604198. DOI: 10.3390/bioengineering10101135.


Inferring light responses of primate retinal ganglion cells using intrinsic electrical signatures.

Zaidi M, Aggarwal G, Shah N, Karniol-Tambour O, Goetz G, Madugula S J Neural Eng. 2023; 20(4).

PMID: 37433293 PMC: 11067857. DOI: 10.1088/1741-2552/ace657.