Albuterol Delivery Via Tracheostomy Tube
Overview
Authors
Affiliations
Hypothesis: Albuterol delivery through a tracheostomy tube is affected by device (nebulizer vs metered-dose inhaler), interface (mask vs T-piece), bias flow, and humidification.
Methods: A lift bar was placed between the chambers of a dual-chambered lung model such that a ventilator triggered simulated spontaneous breathing at a rate of 20 breaths/min, tidal volume of 0.4 L, and inspiratory-expiratory ratio of 1:2. An 8-mm inner diameter cuffed tracheostomy tube was placed through a semi-circular model that simulated a patient's neck. Four conditions of gas flow and humidification were used for the nebulizer experiments: heated aerosol (approximately 30 L/min, approximately 30 degrees C), heated humidity (approximately 30 L/min, approximately 30 degrees C), high flow without added humidity (approximately 30 L/min), or a nebulizer attached to the tracheostomy tube without additional flow. The nebulizer was filled with 4 mL that contained 2.5 mg of albuterol, and operated at 8 L/min. The nebulizer was tested with a T-piece or tracheostomy mask. For the metered-dose inhaler experiments, a spacer was used and actuation of the inhaler (100 microg per actuation) was synchronized with inhalation (4 actuations separated by > or = 15 s). When the spacer was used without additional flow, a valved T-piece was used with a 1-way valve placed either proximal or distal to the spacer. A filter was attached between the lung model and the distal end of the tracheostomy tube. Albuterol washed from the filter was measured by ultraviolet spectrophotometry.
Results: For the nebulizer, the most efficient delivery was with no flow other than that to power the nebulizer and with a T-piece (p < 0.001). The most efficient method for aerosol delivery was metered-dose inhaler with a valved T-piece and placement of the 1-way valve in the proximal position (p < 0.001). The effect of humidity was unclear from the results of this study.
Conclusions: Albuterol delivery via tracheostomy was affected by the delivery device (nebulizer vs inhaler), bias gas flow, and the patient interface.
Albuainain F, Li J Respir Care. 2024; 69(7):847-853.
PMID: 38485144 PMC: 11285491. DOI: 10.4187/respcare.11495.
Application of aerosol therapy in respiratory diseases in children: A Saudi expert consensus.
Alharbi A, Yousef A, Alharbi S, Al-Shamrani A, Alqwaiee M, Almeziny M Ann Thorac Med. 2021; 16(2):188-218.
PMID: 34012486 PMC: 8109687. DOI: 10.4103/atm.atm_74_21.
Ari A, Fink J Can J Respir Ther. 2021; 57:49-52.
PMID: 33959675 PMC: 8086593. DOI: 10.29390/cjrt-2020-054.
Fundamentals of aerosol therapy in critical care.
Dhanani J, Fraser J, Chan H, Rello J, Cohen J, Roberts J Crit Care. 2016; 20(1):269.
PMID: 27716346 PMC: 5054555. DOI: 10.1186/s13054-016-1448-5.