Utility of a Voice-activated System in Minimally Invasive Surgery

Overview
Specialty Gastroenterology
Date 2005 Sep 28
PMID 16185114
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Advances in computer enhancements for surgery, including a voice-activated control system for minimally invasive surgery, are being introduced into clinical practice. Few reports have objectively evaluated the utility of the voice-activated control system. The aim of this study was to evaluate the utility of a voice-activated control system for delivery of commands to specific operating room (OR) equipment.

Materials And Methods: We evaluated a total of 30 laparoscopic procedures: 15 laparoscopic hernia repairs, 10 laparoscopic cholecystectomies, and 5 laparoscopic fundoplications performed with the HERMEStrade mark Operating Room Control Center (Computer Motion, Santa Barbara, California) voice-activated control system. When the voice command (VC) to the HERMES system was given, the circulating nurse was observed and her location was recorded. The 3 locations were A, the nurse was immediately available in the OR, not engaged in any other tasks; B, the nurse was in the OR but engaged in other tasks; or C, the nurse was outside the OR.

Results: The 30 cases were performed by the same surgeon and completed laparoscopically. For 170 VCs for gas insufflation, the location of the circulating nurse was: A=50, B=98, and C=22. For 135 light source adjustments, the locations were A=34, B=82, and C=19. In 76 white balance VCs: A=16, B=51, and C=9. In 128 VCs for camera adjustment: A=27, B=77, and C=24. Eight video capture VCs: A=1, B=3, and C=4. For 69 image capture VCs: A=5, B=41, and C=23. The total number of VCs was 586: A=133 (22.7%), B=352 (60%), and C=101(17.3%). All VCs were accurately interpreted by the HERMES system.

Conclusion: Voice-activated control systems improve communication with and efficiency of OR staff. The surgeon is afforded the most timely equipment adjustment possible. Circulating nurses are allowed to concentrate on patient care instead of equipment adjustment during the course of the surgery.

Citing Articles

Implementation of integrated operating rooms: how much time is saved and how do medical staff experience the upgrading? A mixed methods study in Denmark.

Carstensen K, Jensen E, Madsen M, Thomsen A, Lovschall C, Tayyari Dehbarez N BMJ Open. 2020; 10(7):e034459.

PMID: 32727737 PMC: 7394299. DOI: 10.1136/bmjopen-2019-034459.


Electronic Health Record Interactions through Voice: A Review.

Kumah-Crystal Y, Pirtle C, Whyte H, Goode E, Anders S, Lehmann C Appl Clin Inform. 2018; 9(3):541-552.

PMID: 30040113 PMC: 6051768. DOI: 10.1055/s-0038-1666844.


Touchless interaction with software in interventional radiology and surgery: a systematic literature review.

Mewes A, Hensen B, Wacker F, Hansen C Int J Comput Assist Radiol Surg. 2016; 12(2):291-305.

PMID: 27647327 DOI: 10.1007/s11548-016-1480-6.


Integrated operation systems and voice recognition in minimally invasive surgery: comparison of two systems.

Perrakis A, Hohenberger W, Horbach T Surg Endosc. 2012; 27(2):575-9.

PMID: 22926891 DOI: 10.1007/s00464-012-2488-9.


Equipment and technology in surgical robotics.

Sim H, Yip S, Cheng C World J Urol. 2006; 24(2):128-35.

PMID: 16538515 DOI: 10.1007/s00345-006-0070-6.