A Feasibility Study for Image Guided Radiotherapy Using Low Dose, High Speed, Cone Beam X-ray Volumetric Imaging

Overview
Journal Radiother Oncol
Specialties Oncology
Radiology
Date 2005 Sep 15
PMID 16157400
Citations 29
Authors
Affiliations
Soon will be listed here.
Abstract

Background And Purpose: Image Guidance of patient set-up for radiotherapy can be achieved by acquiring X-ray volumetric images (XVI) with Elekta Synergy and registering these to the planning CT scan. This enables full 3D registration of structures from similar 3D imaging modalities and offers superior image quality, rotational set-up information and a large field of view. This study uses the head section of the Rando phantom to demonstrate a new paradigm of faster, lower dose XVI that still allows registration to high precision.

Materials And Methods: One high exposure XVI scan and one low exposure XVI scan were performed with a Rando Head Phantom. The second scan was used to simulate ultra low dose, fast acquisition, full and half scans by discarding a large number of projections before reconstruction. Dose measurements were performed using Thermo Luminescent Dosimeters (TLD) and an ion chamber. The reconstructed XVI scans were automatically registered with a helical CT scan of the Rando Head using the volumetric, grey-level, cross-correlation algorithm implemented in the Syntegra software package (Philips Medical Systems). Reproducibility of the registration process was investigated.

Results: In both XVI scans the body surface, bone-tissue and tissue air interfaces were clearly visible. Although the subjective image quality of the low dose cone beam scan was reduced, registration of both cone beam scans with the planning CT scan agreed within 0.1 mm and 0.1 degrees . Dose to the patient was reduced from 28mGy to less than 1mGy and the equivalent scan speed reduced to one minute or less.

Conclusions: Automatic 3D registration of high speed, ultra low dose XVI scans with the planning CT scan can be used for precision 3D patient set-up verification/image guidance on a daily basis with out loss of accuracy when compared to higher dose XVI scans.

Citing Articles

Evaluation of a new acrylic-lead shielding device for peripheral dose reduction during cone-beam computed tomography.

Shimizu H, Sasaki K, Aoyama T, Iwata T, Kitagawa T, Kodaira T BJR Open. 2024; 4(1):20220043.

PMID: 38525166 PMC: 10958996. DOI: 10.1259/bjro.20220043.


Evaluation of In-room Volumetric Imaging Doses for Image-guided Radiotherapy: A Multi-institutional Study.

Sakai Y, Monzen H, Takei Y, Kosaka H, Nakamura K, Yanagi Y J Med Phys. 2023; 48(2):189-194.

PMID: 37576099 PMC: 10419753. DOI: 10.4103/jmp.jmp_109_22.


Evaluation of patient organ doses from kilovoltage cone-beam CT imaging in radiation therapy.

Ozseven A, Dirican B Rep Pract Oncol Radiother. 2021; 26(2):251-258.

PMID: 34211776 PMC: 8241296. DOI: 10.5603/RPOR.a2021.0038.


Registration accuracy with the low dose kilovoltage cone-beam CT: A phantom study.

Takei Y, Monzen H, Matsumoto K, Hanaoka K, Tamura M, Nishimura Y BJR Open. 2020; 1(1):20190028.

PMID: 33178952 PMC: 7592495. DOI: 10.1259/bjro.20190028.


Dose reduction potential of using gold fiducial markers for kilovoltage image-guided radiotherapy.

Takei Y, Monzen H, Tamura M, Doi H, Nishimura Y J Appl Clin Med Phys. 2020; 21(10):151-157.

PMID: 32959957 PMC: 7592977. DOI: 10.1002/acm2.13023.