Optimized Scansystem Platelet Kit for Bacterial Detection with Enhanced Sensitivity: Detection Within 24 H After Spiking

Overview
Journal Vox Sang
Specialty Hematology
Date 2005 Sep 9
PMID 16146505
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Background And Objectives: The prevention and detection of bacterial contamination of platelet concentrates remains a major challenge for transfusion medicine. To be suitable for blood-transfusion services, the contamination detection method must be highly sensitive, easy to perform and preferably of low cost. In this spiking study, we evaluated the new optimized Scansystem Platelet Kit detection method for use on apheresis platelets.

Study Design And Methods: Apheresis platelet concentrates (APCs) were individually spiked with 10 colony-forming units (CFU)/ml of one of 10 different strains of bacteria. The spiked APCs were analysed at specific time-points during incubation by using the optimized Scansystem Platelet Kit. Bacterial enumeration was performed by plating onto blood agar.

Results: All the bacterial strains tested were detected by using the optimized Scansystem Platelet Kit when sampled 24 h after spiking. Compared to the Scansystem standard kit, sensitivity was increased to < 50 CFU/ml. The identity of the spiked bacteria was confirmed by Gram staining and DNA fingerprinting.

Conclusion: The optimized Scansystem Platelet Kit was able to reliably detect, within 70 min, 10 transfusion-relevant bacterial species in APCs when a sample volume was taken 24 h after spiking. This is the first study carried out by using the optimized Scansystem bacterial detection that was found to have an enhanced sensitivity compared to the standard kit.

Citing Articles

Diagnostic methods for platelet bacteria screening: current status and developments.

Stormer M, Vollmer T Transfus Med Hemother. 2014; 41(1):19-27.

PMID: 24659944 PMC: 3949613. DOI: 10.1159/000357651.


Impact of donor arm cleaning with different aseptic solutions for prevention of contamination in blood bags.

Patel T, Shukla R, Gupte S Indian J Hematol Blood Transfus. 2014; 29(1):17-20.

PMID: 24426327 PMC: 3572260. DOI: 10.1007/s12288-011-0138-3.


Implementation of Bacterial Detection Methods into Blood Donor Screening - Overview of Different Technologies.

Schmidt M, Sireis W, Seifried E Transfus Med Hemother. 2011; 38(4):259-265.

PMID: 22016697 PMC: 3190223. DOI: 10.1159/000330305.


From the donor's arm to blood product: a study on bacterial contamination of apheresis platelet concentrates.

Leo A, Salvadego M, Piva M, Ruffato G, Valverde S, Trabuio E Blood Transfus. 2009; 5(3):130-3.

PMID: 19204765 PMC: 2535893. DOI: 10.2450/2007.0003-07.


Rapid chromatic detection of bacteria by use of a new biomimetic polymer sensor.

Silbert L, Ben Shlush I, Israel E, Porgador A, Kolusheva S, Jelinek R Appl Environ Microbiol. 2006; 72(11):7339-44.

PMID: 16997991 PMC: 1636204. DOI: 10.1128/AEM.01324-06.