» Articles » PMID: 16103556

Detection of Coronary Microembolisation by Doppler Ultrasound During Percutaneous Coronary Interventions

Overview
Journal Heart
Date 2005 Aug 17
PMID 16103556
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To validate an intracoronary Doppler ultrasound device for high intensity transient signals (HITS) detection and to assess the incidence of HITS during percutaneous coronary intervention (PCI).

Methods And Results: In an in vitro model, particle count and number of HITS detected by an intracoronary 0.014 inch Doppler wire were closely correlated (r = 0.97, p < 0.001). In the clinical study, 32 patients (mean (SD) age 61 (11) years; 23 men, nine women) with coronary artery disease were treated with balloon dilatation and stent implantation for a single vessel stenosis. In these patients HITS were detected during PCI in 84% (27 of 32). Reproducibility (r = 0.99, p < 0.001) and interobserver agreement (r = 0.84, p < 0.001) of HITS counts were significant. The number of HITS after stent implantation was significantly higher than after balloon dilatation (11 (7) v 2 (4), p < 0.001). Postprocedural coronary flow velocity reserve (CFVR) was < 2.0 in 55% (16 of 29) of all patients after balloon dilatation and < 2.0 in 23% (six of 26) after stent implantation. The number of HITS after stent implantation did not differ significantly between patients with CFVR < 2.0 and patients with CFVR > or = 2.0 (12 (8) v 10 (7), not significant).

Conclusions: Embolic particles can be detected as HITS by an intracoronary Doppler ultrasound device. Coronary microembolism is often observed during PCI, especially after stent implantation. However, the incidence of HITS alone does not explain a reduced CFVR after PCI.

Citing Articles

Ischemic Heart Disease Pathophysiology Paradigms Overview: From Plaque Activation to Microvascular Dysfunction.

Severino P, DAmato A, Pucci M, Infusino F, Adamo F, Birtolo L Int J Mol Sci. 2020; 21(21).

PMID: 33143256 PMC: 7663258. DOI: 10.3390/ijms21218118.


Spontaneous and procedural plaque embolisation in native coronary arteries: pathophysiology, diagnosis, and prevention.

De Maria G, Patel N, Kassimis G, Banning A Scientifica (Cairo). 2014; 2013:364247.

PMID: 24455430 PMC: 3881665. DOI: 10.1155/2013/364247.


Cardioprotective Effects of Atorvastatin plus Trimetazidine in Percutaneous Coronary Intervention.

Lin X, Ma A, Zhang W, Lu Q, Sun C, Tian H Pak J Med Sci. 2013; 29(2):545-8.

PMID: 24353574 PMC: 3809233. DOI: 10.12669/pjms.292.2937.


Detection of coronary microembolization by Doppler ultrasound in patients with stable angina pectoris during percutaneous coronary interventions under an adjunctive antithrombotic therapy with abciximab: design and rationale of the High Intensity....

Kretzschmar D, Jung C, Otto S, Utschig S, Hartmann M, Lehmann T Cardiovasc Ultrasound. 2012; 10:21.

PMID: 22613136 PMC: 3407765. DOI: 10.1186/1476-7120-10-21.


[Diagnostics of acute myocardial infarction in elderly patients].

Bahrmann P, Heppner H, Bahrmann A, Christ M, Bertsch T, Sieber C Z Gerontol Geriatr. 2011; 44(3):166-71.

PMID: 21573908 DOI: 10.1007/s00391-011-0196-z.


References
1.
Eicke B, Barth V, Kukowski B, Werner G, Paulus W . Cardiac microembolism: prevalence and clinical outcome. J Neurol Sci. 1996; 136(1-2):143-7. DOI: 10.1016/0022-510x(95)00299-h. View

2.
. Basic identification criteria of Doppler microembolic signals. Consensus Committee of the Ninth International Cerebral Hemodynamic Symposium. Stroke. 1995; 26(6):1123. View

3.
KONG T, Davidson C, Meyers S, Tauke J, Parker M, Bonow R . Prognostic implication of creatine kinase elevation following elective coronary artery interventions. JAMA. 1997; 277(6):461-6. View

4.
Pettelot G, Bracco J, Gibelin P, Baudouy M, Barrillon D, Morand P . Detection of embolic signals using Doppler ultrasound: a new approach to cardiac embolism. Int J Cardiol. 1997; 58(1):1-5. DOI: 10.1016/s0167-5273(96)02851-3. View

5.
Kern M, Dupouy P, Drury J, Aguirre F, Aptecar E, Bach R . Role of coronary artery lumen enlargement in improving coronary blood flow after balloon angioplasty and stenting: a combined intravascular ultrasound Doppler flow and imaging study. J Am Coll Cardiol. 1997; 29(7):1520-7. DOI: 10.1016/s0735-1097(97)00082-x. View