Discrepancies Between Catheter Tip and Tissue Temperature in Cooled-tip Ablation: Relevance to Guiding Left Atrial Ablation

Overview
Journal Circulation
Date 2005 Aug 10
PMID 16087799
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

Background: It is not known whether catheter tip temperatures with a cooled-tip ablation can be reliably extrapolated to estimate actual tissue temperatures. The relationship between catheter tip temperatures, tissue temperatures, power, and microbubble formation is not known.

Methods And Results: Nine dogs underwent 111 radiofrequency energy deliveries at the pulmonary vein ostia with a cooled-tip catheter. Catheter tip and tissue temperatures were markedly discrepant. Catheter tip temperature plateaus at 36 degrees C to 39 degrees C with increasing power, whereas tissue temperature increases to a mean of 75+/-3 degrees C at 45 W (maximum temperature >100 degrees C). Seventy-two energy deliveries were performed, titrating power to microbubble formation guided by intracardiac echocardiography. Type I and II microbubble formation occurred in 45 (63%) and 19 (26%) ablations, respectively. Type I microbubble emergence occurred at lower powers (21+/-8 versus 26+/-4 W; P=0.05), catheter tip temperatures (38+/-5 degrees C versus 48+/-10 degrees C; P=0.02), and tissue temperatures (65+/-19 degrees C versus 81+/-9 degrees C; P<0.001) than type II microbubble formation. Maximum impedance decreases during ablation before microbubble formation were less with type I microbubble (20+/-9 versus 37+/-11 Omega; P<0.001) compared with type II microbubbles. One quarter of type I microbubbles abruptly transitioned to type II microbubbles with significant changes in power or catheter tip temperature. No microbubbles were seen in 19 ablations (26%) despite powers up to 26+/-9 W and tissue temperatures up to 81+/-17 degrees C.

Conclusions: Catheter tip and tissue temperatures are markedly discrepant during cooled-tip ablation. Type I and II microbubble formation occurs at overlapping power and catheter tip and tissue temperature ranges. Neither the absence of microbubbles nor the presence of type I microbubble formation ensures against excessive tissue heating. The appearance of microbubbles may indicate possible tissue overheating and signal a need to decrease energy.

Citing Articles

Evaluation and comparison of impedance and amplitude changes in lesion index-guided pulmonary vein isolation.

Kaneko Y, Naruse Y, Narumi T, Sano M, Urushida T, Maekawa Y J Arrhythm. 2024; 40(1):100-108.

PMID: 38333375 PMC: 10848590. DOI: 10.1002/joa3.12966.


Safe and effective delivery of high-power, short-duration radiofrequency ablation lesions with a flexible-tip ablation catheter.

Ptaszek L, Koruth J, Santangeli P, Piccini J, Ranjan R, Mahapatra S Heart Rhythm O2. 2023; 4(1):42-50.

PMID: 36713045 PMC: 9877396. DOI: 10.1016/j.hroo.2022.10.009.


Temperature-controlled ablation of the mitral isthmus line using the novel DiamondTemp ablation system.

Rottner L, My I, Schleberger R, Moser F, Moser J, Kirchhof P Front Cardiovasc Med. 2022; 9:1046956.

PMID: 36505349 PMC: 9729688. DOI: 10.3389/fcvm.2022.1046956.


Impedance-guided Radiofrequency Ablation: Using Impedance to Improve Ablation Outcomes.

Chinitz J, Michaud G, Stephenson K J Innov Card Rhythm Manag. 2020; 8(10):2868-2873.

PMID: 32477757 PMC: 7252711. DOI: 10.19102/icrm.2017.081003.


Atrial fibrillation ablation using very short duration 50 W ablations and contact force sensing catheters.

Winkle R, Moskovitz R, Hardwin Mead R, Engel G, Kong M, Fleming W J Interv Card Electrophysiol. 2018; 52(1):1-8.

PMID: 29460232 PMC: 5924666. DOI: 10.1007/s10840-018-0322-6.