Visual Simulation of Retinal Images Through a Decentered Monofocal and a Refractive Multifocal Intraocular Lens

Overview
Specialty Ophthalmology
Date 2005 Aug 3
PMID 16075326
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate the effect of decentration of a monofocal intraocular lens (IOL) and a refractive multifocal IOL on retinal image quality using a new visual simulation system.

Methods: Using a new visual simulation system, we performed visual simulation of a monofocal and a multifocal IOL at 5, 4, 3, 2, 1, and 0.4 m with several decentered IOL positions from 0 to 1.0 mm through a 3- or 4-mm aperture using Landolt visual acuity (VA) charts. The VA was estimated under each condition from the simulated retinal image.

Results: With a monofocal IOL, the image was affected minimally by decentration at 4 and 5 m; at 2 and 3 m, the image contrast decreased slightly with increased decentration. With the multifocal IOL, some loss of image contrast developed at all distances compared with the monofocal IOL; however, the images of the Landolt's rings were still recognizable under all conditions.

Conclusions: Our results suggest that up to 1.0 mm of decentration of a monofocal and multifocal IOL would not greatly affect the retinal image quality.

Citing Articles

A Comparative Analysis of the Effects of Misaligning Different Trifocal Intraocular Lenses.

Yan W, Auffarth G, Khoramnia R, Labuz G J Clin Med. 2025; 14(1.

PMID: 39797271 PMC: 11721922. DOI: 10.3390/jcm14010187.


Quantification of ghosting produced with presbyopic contact lens correction.

Kollbaum P, Dietmeier B, Jansen M, Rickert M Eye Contact Lens. 2012; 38(4):252-9.

PMID: 22735627 PMC: 3713236. DOI: 10.1097/ICL.0b013e31825aa879.


Influence of tilt and decentration of scleral-sutured intraocular lens on ocular higher-order wavefront aberration.

Oshika T, Sugita G, Miyata K, Tokunaga T, Samejima T, Okamoto C Br J Ophthalmol. 2006; 91(2):185-8.

PMID: 16914469 PMC: 1857623. DOI: 10.1136/bjo.2006.099945.

References
1.
Negishi K, Bissen-Miyajima H, Kato K, Kurosaka D, Nagamoto T . Evaluation of a zonal-progressive multifocal intraocular lens. Am J Ophthalmol. 1998; 124(3):321-30. DOI: 10.1016/s0002-9394(14)70824-4. View

2.
Portney V . Optical testing and inspection methodology for modern intraocular lenses. J Cataract Refract Surg. 1992; 18(6):607-13. DOI: 10.1016/s0886-3350(13)80453-1. View

3.
Ravalico G, Baccara F, Bellavitis A . Refractive bifocal intraocular lens and pupillary diameter. J Cataract Refract Surg. 1992; 18(6):594-7. DOI: 10.1016/s0886-3350(13)80450-6. View

4.
CAMPBELL F, Green D . Optical and retinal factors affecting visual resolution. J Physiol. 1965; 181(3):576-93. PMC: 1357668. DOI: 10.1113/jphysiol.1965.sp007784. View

5.
Korynta J, Bok J, Cendelin J . Changes in refraction induced by change in intraocular lens position. J Refract Corneal Surg. 1994; 10(5):556-64. View