» Articles » PMID: 16060445

Parallel Comparison of Grip Strength Measures Obtained with a MicroFET 4 and a Jamar Dynamometer

Overview
Date 2005 Aug 3
PMID 16060445
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Repeated measures of grip strength obtained bilaterally with a Jamar and a MicroFET 4 dynamometer were compared. Measurements obtained with the MicroFET 4 tended to be slightly (2.2-3.1 lb.) higher but were highly correlated (r > or = .96) with those obtained with the Jamar. Parallel reliability for the two devices was excellent (intraclass correlation coefficient > or = .96). Although clinicians should be cautious about using the devices interchangeably, the MicroFET 4 appears to be a legitimate alternative to the Jamar dynamometer.

Citing Articles

Variability between Different Hand-Held Dynamometers for Measuring Muscle Strength.

Du W, Cornett K, Donlevy G, Burns J, McKay M Sensors (Basel). 2024; 24(6).

PMID: 38544123 PMC: 10974287. DOI: 10.3390/s24061861.


Gripwise Versus Jamar: The Challenge of a New Dynamometer Assessing Handgrip Strength.

Villain C, Lebaube S, Kremer C, Chavoix C, Fournel F, Briant A J Gerontol A Biol Sci Med Sci. 2023; 78(12):2458-2465.

PMID: 37578949 PMC: 10692423. DOI: 10.1093/gerona/glad198.


Measures of Maximal Tactile Pressures during a Sustained Grasp Task Using a TactArray Device Have Satisfactory Reliability and Concurrent Validity in People with Stroke.

Gopaul U, Laver D, Carey L, Matyas T, van Vliet P, Callister R Sensors (Basel). 2023; 23(6).

PMID: 36992002 PMC: 10059963. DOI: 10.3390/s23063291.


Physical activity and fitness in survivors of head and neck cancer.

Fang Y, Wang C, Chen Y, Lou P, Ko J, Lin J Support Care Cancer. 2021; 29(11):6807-6817.

PMID: 33997941 DOI: 10.1007/s00520-021-06192-y.


Grip strength measured by high precision dynamometry in healthy subjects from 5 to 80 years.

Hogrel J BMC Musculoskelet Disord. 2015; 16:139.

PMID: 26055647 PMC: 4460675. DOI: 10.1186/s12891-015-0612-4.