» Articles » PMID: 16023817

Fluorescence Detection Combined with Either HPLC or HPTLC for Pharmaceutical Quality Control in a Hospital Chemotherapy Production Unit: Application to Camptothecin Derivatives

Overview
Specialty Chemistry
Date 2005 Jul 19
PMID 16023817
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

In order to achieve the analytical assessment of the manufactured batches of chemotherapy preparations, post-production quality control has been developed. The common use of camptothecin derivatives (i.e. irinotecan (CPT-11) and topotecan (TPT)) as part of protocols in Institut Gustave Roussy (IGR) has led to develop an efficient analytical method that could assess an increasing number of samples with high throughput, good specificity and practicality. Due to the difference of concentration between batches containing irinotecan or topotecan, HPLC and HPTLC both combined with fluorescence detection were investigated. Those two techniques made identity, purity and quantitation assays possible. The chromatographic conditions that were chosen allowed identification of each drug through their rate of flow (Rf), 0.10 and 0.35, or their retention time (tR), 2 and 7 min for topotecan and irinotecan, respectively. A calibration curve was plotted for each molecule and validated by three quality controls (high, medium and low). Coefficients of variation of repeatability (CVr) and intermediate precision (CVi) were determined for both methods. Considering their values and the concentration ranges (from 100 to 500 mg/L for HPTLC and from 0.1 to 1 mg/L for HPLC), it was decided to perform analysis using HPTLC for irinotecan preparations and HPLC for topotecan preparations. These inferences seemed appropriate regarding the number of preparations to be assayed.

Citing Articles

Poly(amidoamine) Dendrimer/Camptothecin Complex: From Synthesis to In Vitro Cancer Cell Line Studies.

Oledzka E, Pasnik K, Domanska I, Zielinska-Pisklak M, Piotrowska U, Sobczak M Molecules. 2023; 28(6).

PMID: 36985668 PMC: 10052527. DOI: 10.3390/molecules28062696.


Development and validation of an UPLC-MS/MS method for the quantification of irinotecan, SN-38 and SN-38 glucuronide in plasma, urine, feces, liver and kidney: Application to a pharmacokinetic study of irinotecan in rats.

Basu S, Zeng M, Yin T, Gao S, Hu M J Chromatogr B Analyt Technol Biomed Life Sci. 2016; 1015-1016:34-41.

PMID: 26894853 PMC: 5215916. DOI: 10.1016/j.jchromb.2016.02.012.