How Well Does Your Ruby Laser Work?
Overview
Affiliations
The safe and effective use of cutaneous lasers is dependent upon the selection of a clinically appropriate laser energy level. Therefore, the accuracy of the stated laser output is critical. The energy output of an Aesculap ruby laser was recorded at the start of every clinical session over a 704-day period using an external energy meter. Regular review of these records enabled trends of energy output to be recognised and appropriate action to be taken prior to adverse clinical sequelae. This monitoring procedure is a quick and inexpensive method of improving quality control in the use of medical lasers.
Ash C, Donne K, Daniel G, Town G, Clement M, Valentine R J Appl Clin Med Phys. 2012; 13(5):3702.
PMID: 22955640 PMC: 5718238. DOI: 10.1120/jacmp.v13i5.3702.
Adverse effects reported in epilatory ruby laser treatment.
Royston S, Wright P, Widdowson D, Wareham W, Strike P Lasers Med Sci. 2007; 23(1):35-9.
PMID: 17431730 DOI: 10.1007/s10103-007-0451-0.
Monitoring the ups and downs of pulsed dye laser energy output.
Wright P, Widdowson D, Ahmed S, Moore J Lasers Med Sci. 2006; 21(1):49-53.
PMID: 16583184 DOI: 10.1007/s10103-006-0372-3.