Electromagnetic Interference of Implantable Unipolar Cardiac Pacemakers by an Induction Oven

Overview
Date 2005 Jun 16
PMID 15955187
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Induction ovens have been reported to exert electromagnetic interference on implanted cardiac pacemakers. In an attempt to quantitatively investigate the electromagnetic interference caused by an induction oven on implantable unipolar cardiac pacemakers, we measured the distribution profile of the magnetic field intensity, both with and without a pan on the induction oven. We also performed the inhibition test and asynchronous test using four kinds of pacemakers housed in the standardized Irnich human body model, and measured the maximal distance from the induction oven up to which the interference occurred. In the pan-detection mode of the oven in the absence of a pan, the distribution profile of the magnetic field intensity peaked at the center of the cooking plate, and during induction heating of a pan placed on the induction oven, it was the largest at the circular top-edge of the pan. Pacemaker pulses were inhibited by the induction oven, or generated by the reversion mechanism. The maximal interference distance from the oven was 34 cm for one of the pacemakers. Thus, the safe distance from an induction oven of a patient with an implanted cardiac pacemaker is considered to be 50 cm or more. In conclusion, in the pan-detection mode of the oven in the absence of a pan, the distribution profile of the magnetic field intensity peaked at the center of the cooking plate, and during the induction heating of a pan placed on the oven, it peaked at the circular edge of the pan. The induction oven asynchronized or generated pulses in implantable unipolar cardiac pacemakers up to a maximal distance of 34 cm from the induction oven.

Citing Articles

Interference-free nanogap pressure sensor array with high spatial resolution for wireless human-machine interfaces applications.

Yang J, Chung M, Yoo J, Kim M, Kim B, Jo M Nat Commun. 2025; 16(1):2024.

PMID: 40016221 PMC: 11868543. DOI: 10.1038/s41467-025-57232-8.


Electromagnetic interference in cardiac electronic implants caused by novel electrical appliances emitting electromagnetic fields in the intermediate frequency range: a systematic review.

Driessen S, Napp A, Schmiedchen K, Kraus T, Stunder D Europace. 2018; 21(2):219-229.

PMID: 29992289 PMC: 6365808. DOI: 10.1093/europace/euy155.


Electromagnetic interference of implantable cardiac devices from a shoulder massage machine.

Yoshida S, Fujiwara K, Kohira S, Hirose M J Artif Organs. 2014; 17(3):243-9.

PMID: 24710851 DOI: 10.1007/s10047-014-0765-1.


Mechanized scaling with ultrasonics: Perils and proactive measures.

Paramashivaiah R, Prabhuji M J Indian Soc Periodontol. 2013; 17(4):423-8.

PMID: 24174718 PMC: 3800401. DOI: 10.4103/0972-124X.118310.


Testing of common electromagnetic environments for risk of interference with cardiac pacemaker function.

Tiikkaja M, Aro A, Alanko T, Lindholm H, Sistonen H, Hartikainen J Saf Health Work. 2013; 4(3):156-9.

PMID: 24106646 PMC: 3791084. DOI: 10.1016/j.shaw.2013.06.002.