Evaluation of a Continuous Glucose Monitoring System in Cats with Diabetes Mellitus

Overview
Date 2005 Jun 1
PMID 15922222
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

A continuous glucose monitoring system (CGMS) was evaluated in 14 cats with naturally occurring diabetes mellitus. The device measures interstitial fluid glucose continuously, by means of a sensor placed in the subcutaneous tissue. All cats tolerated the device well and a trace was obtained on 15/16 occasions. There was good correlation between the CGMS values and blood glucose concentration measured using a glucometer (r=0.932, P<0.01). Limitations to the use of the CGMS are its working glucose range of 2.2-22.2 mmol/l (40-400 mg/dl) and the need for calibration with a blood glucose measurement at least every 12 h. When compared to a traditional blood glucose curve, the CGMS is minimally invasive, reduces the number of venepunctures necessary to assess the kinetics of insulin therapy in a patient and provides a truly continuous glucose curve.

Citing Articles

Comparison between a Flash Glucose Monitoring System and a Portable Blood Glucose Meter for Monitoring of Cats with Diabetic Ketosis or Ketoacidosis.

Eiermann J, Hazuchova K, Vidondo B, Campos M, Schuller S, Cui Y Animals (Basel). 2024; 14(19).

PMID: 39409736 PMC: 11475600. DOI: 10.3390/ani14192787.


Evaluation of the FreeStyle Libre, a flash glucose monitoring system, in client-owned cats with diabetes mellitus.

Knies M, Teske E, Kooistra H J Feline Med Surg. 2022; 24(8):e223-e231.

PMID: 35762266 PMC: 9315169. DOI: 10.1177/1098612X221104051.


Validation of a flash glucose monitoring system in outpatient diabetic cats.

Shea E, Hess R J Vet Intern Med. 2021; 35(4):1703-1712.

PMID: 34223655 PMC: 8295694. DOI: 10.1111/jvim.16216.


Accuracy of a flash glucose monitoring system in cats and determination of the time lag between blood glucose and interstitial glucose concentrations.

Baldo F, Fracassi F, Pires J, Tardo A, Malerba E, Manassero E J Vet Intern Med. 2021; 35(3):1279-1287.

PMID: 33991129 PMC: 8163110. DOI: 10.1111/jvim.16122.


Accuracy of a flash glucose monitoring system in dogs with diabetic ketoacidosis.

Malerba E, Cattani C, Baldo F, Carotenuto G, Corradini S, Golinelli S J Vet Intern Med. 2019; 34(1):83-91.

PMID: 31725202 PMC: 6979271. DOI: 10.1111/jvim.15657.


References
1.
Whitley N, Drobatz K, Panciera D . Insulin overdose in dogs and cats: 28 cases (1986-1993). J Am Vet Med Assoc. 1997; 211(3):326-30. View

2.
Rebrin K, Steil G . Can interstitial glucose assessment replace blood glucose measurements?. Diabetes Technol Ther. 2001; 2(3):461-72. DOI: 10.1089/15209150050194332. View

3.
Bode B, Gross T, Thornton K, Mastrototaro J . Continuous glucose monitoring used to adjust diabetes therapy improves glycosylated hemoglobin: a pilot study. Diabetes Res Clin Pract. 2000; 46(3):183-90. DOI: 10.1016/s0168-8227(99)00113-8. View

4.
Goossens M, Nelson R, Feldman E, Griffey S . Response to insulin treatment and survival in 104 cats with diabetes mellitus (1985-1995). J Vet Intern Med. 1998; 12(1):1-6. DOI: 10.1111/j.1939-1676.1998.tb00489.x. View

5.
Monsod T, Flanagan D, Rife F, Saenz R, Caprio S, Sherwin R . Do sensor glucose levels accurately predict plasma glucose concentrations during hypoglycemia and hyperinsulinemia?. Diabetes Care. 2002; 25(5):889-93. DOI: 10.2337/diacare.25.5.889. View