The Effect of Infection-control Barriers on the Light Intensity of Light-cure Units and Depth of Cure of Composite
Overview
Affiliations
Aim: Dental curing lights are vulnerable to contamination with oral fluids during routine intra-oral use. This controlled study aimed to evaluate whether or not disposable transparent barriers placed over the light-guide tip would affect light output intensity or the subsequent depth of cure of a composite restoration.
Methods: The impact on light intensity emitted from high-, medium- and low-output light-cure units in the presence of two commercially available disposable infection-control barriers was evaluated against a no-barrier control. Power density measurements from the three intensity light-cure units were recorded with a radiometer, then converted to a digital image using an intra-oral camera and values determined using a commercial computer program. For each curing unit, the measurements were repeated on ten separate occasions with each barrier and the control. Depth of cure was evaluated using a scrape test in a natural tooth model.
Results: At each level of light output, the two disposable barriers produced a significant reduction in the mean power density readings compared to the no-barrier control (P<0.005). The cure sleeve inhibited light output to a greater extent than either the cling film or the control (P<0.005). Only composite restorations light-activated by the high level unit demonstrated a small but significant decrease in the depth of cure compared to the control (P<0.05).
Conclusion: Placing disposable barriers over the light-guide tip reduced the light intensity from all three curing lights. There was no impact on depth of cure except for the high-output light, where a small decrease in cure depth was noted but this was not considered clinically significant. Disposable barriers can be recommended for use with light-cure lights.
Effect of infection control barriers on the light output from a multi-peak light curing unit.
Soares C, Braga S, Ribeiro M, Price R J Dent. 2020; 103:103503.
PMID: 33091553 PMC: 7569381. DOI: 10.1016/j.jdent.2020.103503.
Infection Control Barrier and Curing Time as Factors Affecting the Irradiance of Light-Cure Units.
Ajaj R, Nassar H, Hasanain F J Int Soc Prev Community Dent. 2019; 8(6):523-528.
PMID: 30596043 PMC: 6280572. DOI: 10.4103/jispcd.JISPCD_226_18.
Evaluation of light-curing units in rural and urban areas.
Alshaafi M Saudi Dent J. 2013; 24(3-4):163-7.
PMID: 23960546 PMC: 3729293. DOI: 10.1016/j.sdentj.2012.07.001.