RF Artifacts Caused by Metallic Implants or Instruments Which Get More Prominent at 3 T: an in Vitro Study

Overview
Publisher Elsevier
Specialty Radiology
Date 2005 May 3
PMID 15862651
Citations 19
Authors
Affiliations
Soon will be listed here.
Abstract

Metallic devices with high electrical conductivity inside or adjacent to the body might lead to marked alterations of the RF amplitude B1 in the tissue under investigation, especially at increased RF frequency, and if specific conditions for electromagnetic resonance are fulfilled. RF-metal interaction effects were investigated systematically at B0=0.2, 1.5 and 3 T analyzing correlated image artifacts for copper wires (d=1 mm, L=53 and 27 cm), and for following instruments and implants made of titanium or nitinol: biopsy needles, hip prostheses, vascular stents and aneurysm clips. The samples were examined in Gd-DTPA-doped 140 mM NaCl solution using spin-echo (SE) sequences with high readout bandwidth. Automatic transmitter adjustment V(T,auto) and manually reduced transmitter voltage VT were applied in order to detect B1 enhancement. At 0.2 T, beyond the shielding of the luminal region of the stents, no RF effects were observed. At 1.5 T, the copper wires caused distinct RF artifacts. At 3 T, RF artifacts also appeared for the hip prostheses and the biopsy needles. Stents with pronounced luminal shielding at lower field strength revealed marked B1 enhancement close to their outer surface.

Citing Articles

Customized Additive Manufacturing in Bone Scaffolds-The Gateway to Precise Bone Defect Treatment.

Zhou J, See C, Sreenivasamurthy S, Zhu D Research (Wash D C). 2023; 6:0239.

PMID: 37818034 PMC: 10561823. DOI: 10.34133/research.0239.


Artifact characterization of Nitinol needles in magnetic resonance imaging-guided musculoskeletal interventions at 3.0 tesla: a phantom study.

Schmidt V, Dietrich O, Seidensticker M, Wildgruber M, Erber B, Ricke J Diagn Interv Radiol. 2023; 29(5):719-732.

PMID: 37592882 PMC: 10679549. DOI: 10.4274/dir.2023.232262.


Magnetic Resonance Visibility, Artifacts, and Overall Safety of the Self-Locating Peritoneal Dialysis Catheter with a Tungsten Tip.

Gallieni M, Rossi U, Tonolini M, Ianniello A, Miglio R, Sabiu G Int J Nephrol. 2023; 2023:7901413.

PMID: 36733472 PMC: 9889163. DOI: 10.1155/2023/7901413.


Artifact reduction of coaxial needles in magnetic resonance imaging-guided abdominal interventions at 1.5 T: a phantom study.

Schmidt V, Arnone F, Dietrich O, Seidensticker M, Armbruster M, Ricke J Sci Rep. 2021; 11(1):22963.

PMID: 34824361 PMC: 8617285. DOI: 10.1038/s41598-021-02434-5.


Susceptibility artifacts from metallic markers and cardiac catheterization devices on a high-performance 0.55 T MRI system.

Basar B, Sonmez M, Yildirim D, Paul R, Herzka D, Kocaturk O Magn Reson Imaging. 2020; 77:14-20.

PMID: 33309924 PMC: 8132593. DOI: 10.1016/j.mri.2020.12.002.