» Articles » PMID: 15829215

A Comparison of Plates with and Without Locking Screws in a Calcaneal Fracture Model

Overview
Journal Foot Ankle Int
Publisher Sage Publications
Specialty Orthopedics
Date 2005 Apr 15
PMID 15829215
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

Background: We compared different plates in an experimental calcaneal fracture model under biocompatible loading.

Methods: Four plates were tested: a plate without locked screws (Synthes), and three different plates with locked screws (Newdeal, Darco, Synthes). Synthetic calcanei (Sawbone) were osteotomized to create a fracture model, and the plates were fixed onto them. Seven specimens for each plate model were subjected to cyclic loading (preload 20 N, 1,000 cycles with 800 N, 0.75 mm/s), and load to failure (0.75 mm/s). Motion, forces, plastic deformation of the plate, and consequent depression of the posterior joint facet were analyzed.

Results: During cyclic loading, all plates with locked screws showed statistically significant lower displacement in the primary loading direction than the plates without locked screws. Mean values (mm) of maximal displacements for each plate during cyclic loading were as follows: Synthes, 3.5; Darco, 4.5; Newdeal, 5.0; Synthes without locked screws, 7.5; (p < 0.001). No statistically significant differences between the plates were found in relation to loads to failure and corresponding displacement.

Conclusion: This is the first biomechanical study to assess the stability of different plates currently in use in our practice for the fixation of calcaneal fractures. Our results showed that plates with locked screws provided greater stability during cyclic loading than the plate without locked screws.

Citing Articles

A 28-Year-Old Woman with Down Syndrome, Congenital Heart Disease, and a History of Knee Surgery and Plantar Fasciitis, with Hallux Abducto Valgus (Bunion) and Lapiplasty Three-Dimensional Correction Surgery.

Simon-Perez E, Jimenez-Martin R, Cicchinelli L, Fernandez Yague J, Simon-Perez C, Paez-Moguer J Am J Case Rep. 2023; 24:e940879.

PMID: 38091276 PMC: 10728881. DOI: 10.12659/AJCR.940879.


[Effectiveness analysis of modified tarsal sinus approach for Sanders - type calcaneal fractures].

Fang K, Ke Q, Wu S, Cai L, Zhang X Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2019; 33(4):462-466.

PMID: 30983195 PMC: 8337175. DOI: 10.7507/1002-1892.201810010.


Sanders type 3-4 calcaneal fractures fixed with locking plate: does tri-cortical iliac bone autograft provide extra benefit for stabilization?.

Bulut T, Gursoy M, Ozturk T, Ozcan C, Sener M Ther Clin Risk Manag. 2018; 14:1665-1670.

PMID: 30254447 PMC: 6140723. DOI: 10.2147/TCRM.S146530.


Radiological study of the secondary reduction effect of early functional exercise on displaced intra-articular calcaneal fractures after internal compression fixation.

Chen W, Liu B, Lv H, Su Y, Chen X, Zhu Y Int Orthop. 2017; 41(9):1953-1961.

PMID: 28660328 DOI: 10.1007/s00264-017-3533-z.


Biomechanical evaluation of reconstruction plates with locking, nonlocking, and hybrid screws configurations in calcaneal fracture: a finite element model study.

Chen C, Hung C, Hsu Y, Chen C, Chiang C Med Biol Eng Comput. 2017; 55(10):1799-1807.