» Articles » PMID: 15798340

Modelling Simple Helically Delivered Dose Distributions

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 2005 Mar 31
PMID 15798340
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

In a previous paper, we described quality assurance procedures for Hi-Art helical tomotherapy machines. Here, we develop further some ideas discussed briefly in that paper. Simple helically generated dose distributions are modelled, and relationships between these dose distributions and underlying characteristics of Hi-Art treatment systems are elucidated. In particular, we describe the dependence of dose levels along the central axis of a cylinder aligned coaxially with a Hi-Art machine on fan beam width, couch velocity and helical delivery lengths. The impact on these dose levels of angular variations in gantry speed or output per linear accelerator pulse is also explored.

Citing Articles

A quality assurance tool for helical tomotherapy using a step-wedge phantom and the on-board MVCT detector.

Althof V, van Haaren P, Westendorp R, Nuver T, Kramer D, Ikink M J Appl Clin Med Phys. 2012; 13(1):3585.

PMID: 22231210 PMC: 5716125. DOI: 10.1120/jacmp.v13i1.3585.


On the impact of longitudinal breathing motion randomness for tomotherapy delivery.

Kissick M, Flynn R, Westerly D, Hoban P, Mo X, Soisson E Phys Med Biol. 2008; 53(18):4855-73.

PMID: 18711250 PMC: 2610272. DOI: 10.1088/0031-9155/53/18/001.


Whole brain radiotherapy with hippocampal avoidance and simultaneously integrated brain metastases boost: a planning study.

Gutierrez A, Westerly D, Tome W, Jaradat H, Mackie T, Bentzen S Int J Radiat Oncol Biol Phys. 2007; 69(2):589-97.