Origins of the Edge Shadowing Artefact in Medical Ultrasound Imaging

Overview
Specialty Radiology
Date 2004 Nov 20
PMID 15550319
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Ultrasound (US) B-mode images distal to smooth, rounded cavities, such as cysts, containing a fluid with a speed of sound mismatch to the surrounding tissue, often exhibit a "refractile" edge shadowing artefact. This usually appears as narrow, hypoechoic, shadow lines extending a significant distance distal to the lateral edges of the fluid cavity and parallel to the US beam. The true reasons for this artefact are likely to be complex and to vary from case to case, with many different explanations found in the literature. However, we present a simplified theoretical model for the phenomenon based on a pulsed, finite-beam solution of US scattering from circular fluid-filled cylinders that suggests that "edge" shadows can occur distal not only to edges but also to points where the incident beam intersects the cavity near to the critical angle. Both mechanisms support the view that edge shadows can arise from the combination of unusually high wavefront spreading and the speckle-generating nature of the surrounding tissue. In vitro data from a tissue-mimicking phantom confirms that the edge shadow structure depends on the sign of the speed of sound contrast between the cylinder fluid and the surrounding medium.

Citing Articles

A MODEL TO PREDICT DIAGNOSIS OF PANCREATIC NEUROENDOCRINE TUMORS BASED ON EUS IMAGING FEATURES.

Saizu I, Cotruta B, Iacob R, Bunduc S, Saizu R, Dumbrava M Acta Endocrinol (Buchar). 2024; 19(4):407-414.

PMID: 38933251 PMC: 11197819. DOI: 10.4183/aeb.2023.407.


B-Mode and Contrast-Enhanced Ultrasonography Aspects of Benign and Malignant Superficial Neoplasms in Dogs: A Preliminary Study.

Hillaert A, Stock E, Duchateau L, de Rooster H, Devriendt N, Vanderperren K Animals (Basel). 2022; 12(20).

PMID: 36290151 PMC: 9597709. DOI: 10.3390/ani12202765.


Predictive value and prevalence of refractive edge shadow in diagnosis of ovarian dermoids.

Shen L, Tse J, Negrete L, Lo E, Yoon L, Kamaya A Abdom Radiol (NY). 2022; 47(12):4227-4236.

PMID: 36098759 DOI: 10.1007/s00261-022-03666-3.


Refraction artifact on abdominal sonogram.

Naganuma H, Ishida H, Uno A, Nagai H, Ogawa M, Kamiyama N J Med Ultrason (2001). 2021; 48(3):273-283.

PMID: 34021821 DOI: 10.1007/s10396-021-01097-z.


Weakly Supervised Estimation of Shadow Confidence Maps in Fetal Ultrasound Imaging.

Meng Q, Sinclair M, Zimmer V, Hou B, Rajchl M, Toussaint N IEEE Trans Med Imaging. 2019; 38(12):2755-2767.

PMID: 31021795 PMC: 6892638. DOI: 10.1109/TMI.2019.2913311.