» Articles » PMID: 15454721

Posterior Augmentation of an Anterior Lumbar Interbody Fusion: Minimally Invasive Fixation Versus Pedicle Screws in Vitro

Overview
Specialty Orthopedics
Date 2004 Sep 30
PMID 15454721
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Study Design: An in vitro biomechanical comparison of four posterior fixation techniques in the setting of an anterior lumbar interbody fusion (ALIF).

Objective: To compare the initial stability, in terms of range of motion and neutral zone, provided by pedicle screws, facet screws, translaminar facet screws, and H-graft plus interspinous cables in the presence of an anteriorly placed femoral ring allograft.

Summary Of Background Data: Pedicular fixation has been used to increase ALIF fusion rates but has also been linked with increased morbidity. Alternative posterior fixation options are available, but comprehensive biomechanical comparisons of these techniques do not exist. METHODS.: Twelve cadaveric lumbar motion segments were loaded to 5 Nm in unconstrained flexion-extension, lateral bending, and axial torsion. Specimens were tested intact, after ALIF, and after applying pedicle screws, translaminar screws, facet screws, and H-graft plus cables. The resulting neutral zones and ranges of motion were measured.

Results: The mean (+/-SEM) range of motion for each construct in flexion-extension was as follows: intact: 6.39 degrees (+/-0.47 degrees); ALIF alone: 3.31 degrees (+/-0.22 degrees); (ALIF+) pedicle screws: 0.6 degrees (+/-0.06 degrees); facet screws: 0.75 degrees (+/-0.12 degrees); translaminar screws: 0.61 degrees (+/-0.09 degrees); and H-graft: 1.74 degrees (+/-0.26 degrees). Pedicle, translaminar facet, and facet screws significantly decreased range of motion and neutral zone compared to ALIF alone in flexion-extension, lateral bending, and axial torsion (all at P < 0.04, except translaminar screws in torsion neutral zone where P = 0.09). H-graft decreasedflexion-extension range of motion and neutral zone only (P < 0.01) and resulted in a significantly greater neutral zone than pedicle and facet screws in torsion and lateral bending neutral zones (P < 0.03).

Conclusions: In the ALIF setting, facet screw and translaminar screw techniques, which may be associated with less morbidity than pedicle screws clinically, provided initial posterior stabilization similar to pedicular fixation in this in vitro study.

Citing Articles

Value of single-level circumferential fusion: a 10-year prospective outcomes and cost-effectiveness analysis comparing posterior facet versus pedicle screw fixation.

Buttermann G, Hollmann S, Arpino J, Ferko N Eur Spine J. 2019; 29(2):360-373.

PMID: 31583439 DOI: 10.1007/s00586-019-06165-0.


A Comparison Study of Three Posterior Fixation Strategies in Transforaminal Lumbar Interbody Fusion Lumbar for the Treatment of Degenerative Diseases.

Hu Y, Zhu B, Kepler C, Yuan Z, Dong W, Sun X Indian J Orthop. 2019; 53(4):542-547.

PMID: 31303670 PMC: 6590021. DOI: 10.4103/ortho.IJOrtho_282_18.


Biomechanical characteristics of an integrated lumbar interbody fusion device.

Voronov L, Vastardis G, Zelenakova J, Carandang G, Havey R, Waldorff E Int J Spine Surg. 2015; 8.

PMID: 25694931 PMC: 4325499. DOI: 10.14444/1001.


Biomechanical evaluation of the Facet Wedge: a refined technique for facet fixation.

Hartensuer R, Riesenbeck O, Schulze M, Gehweiler D, Raschke M, Pavlov P Eur Spine J. 2014; 23(11):2321-9.

PMID: 25155837 DOI: 10.1007/s00586-014-3533-2.


Transforaminal lumbar interbody fusion using unilateral pedicle screw fixation plus contralateral translaminar facet screw fixation in lumbar degenerative diseases.

Liu F, Jiang C, Cao Y, Jiang X, Feng Z Indian J Orthop. 2014; 48(4):374-9.

PMID: 25143640 PMC: 4137514. DOI: 10.4103/0019-5413.136240.