» Articles » PMID: 15334590

Comparison Between 2D and 3D Gradient-echo Sequences for MRI of Human Lung Ventilation with Hyperpolarized 3He

Overview
Journal Magn Reson Med
Publisher Wiley
Specialty Radiology
Date 2004 Aug 31
PMID 15334590
Citations 21
Authors
Affiliations
Soon will be listed here.
Abstract

Images of hyperpolarized 3He were acquired during breath-hold in four healthy volunteers with the use of an optimized 3D gradient-echo sequence. The images were compared with existing 2D gradient-echo methods. The average SNR from a 13-mm-thick slice in the peripheral lung was 1.4 times greater with 3D. In the airways the average SNR was 1.7 times greater with 3D. The higher SNR of 3D was particularly evident when regions of unimpeded gas diffusion, such as the major airways, were imaged with thin slices. This is because diffusion dephasing due to the slice-encoding gradient is minimized with a 3D sequence. The in vivo experimental findings were substantiated with experiments on phantoms of free gas, which showed more than four times the SNR with 3D compared to 2D. Theoretical simulations of the 2D and 3D k-space filters were also performed to predict the SNR and spatial resolution observed in the experimental images.

Citing Articles

Acquiring Hyperpolarized 129Xe Magnetic Resonance Images of Lung Ventilation.

Garrison W, Mugler 3rd J, Mata J, Nunoo-Asare R, Shim Y, Miller G J Vis Exp. 2023; (201).

PMID: 38078603 PMC: 11479144. DOI: 10.3791/65982.


Diffusion weighted hyperpolarized Xe MRI of the lung with 2D and 3D (FLORET) spiral.

Bdaiwi A, Willmering M, Wang H, Cleveland Z Magn Reson Med. 2022; 89(4):1342-1356.

PMID: 36352793 PMC: 9892235. DOI: 10.1002/mrm.29518.


In vivo methods and applications of xenon-129 magnetic resonance.

Marshall H, Stewart N, Chan H, Rao M, Norquay G, Wild J Prog Nucl Magn Reson Spectrosc. 2021; 122:42-62.

PMID: 33632417 PMC: 7933823. DOI: 10.1016/j.pnmrs.2020.11.002.


A multi spin echo pulse sequence with optimized excitation pulses and a 3D cone readout for hyperpolarized C imaging.

Somai V, Wright A, Fala M, Hesse F, Brindle K Magn Reson Med. 2020; 84(4):1895-1908.

PMID: 32173908 PMC: 8638674. DOI: 10.1002/mrm.28248.


He diffusion MRI in human lungs.

Woods J, Conradi M J Magn Reson. 2018; 292:90-98.