» Articles » PMID: 15324756

Mechanism of Oxygen-induced EPR Line Broadening in Lithium Phthalocyanine Microcrystals

Overview
Journal J Magn Reson
Publisher Elsevier
Date 2004 Aug 25
PMID 15324756
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

EPR oximetry has been recognized as an important tool for determining oxygen concentration in biological tissues, in vivo. The method relies on the use of oxygen-sensitive paramagnetic probes whose linewidth varies predictably, mostly linear, with oxygen concentration. Lithium phthalocyanine (LiPc) radical has emerged as the probe of choice due to its superior EPR sensitivity, oxygen response, and biocompatibility. However, there are certain limitations in the preparation of this material in a pure and usable form. In our efforts to improve the synthesis of this material for reliable use in oximetry applications, we developed microcrystalline particulates that showed several advantages over other probes. Despite its advantages, the probe shows linear response to pO2 only in the range of 0-70 mmHg, beyond which a saturation behavior is observed. The goal of this study was to understand the mechanism of the interaction of oxygen with LiPc in order to interpret the experimentally observed linewidths. We propose a dual-spin model in which the freely diffusing spins of LiPc are converted to fixed spins by adsorption of molecular oxygen. The proposed mechanism was verified from the effect of oxygenation/deoxygenation processes on the linewidth of LiPc. In summary, we demonstrated that adsorption of oxygen molecules on LiPc contributes to a nonlinear line-broadening effect. This understanding is important for the future design of new EPR oximetry probes.

Citing Articles

Lithium phthalocyanine (γ-structure) as a molecular oxygen sensor.

Tomsik E, Cernochova Z, Scheibe M, Tadyszak K RSC Adv. 2025; 15(5):3738-3748.

PMID: 39911542 PMC: 11795808. DOI: 10.1039/d4ra08335k.


Composite spin probes with adjustable oxygen sensitivity for pulse electron paramagnetic resonance imaging.

Canavesi I, Viswakarma N, Khurana R, Epel B, Kuppusamy P, Pagel M Magn Reson Med. 2025; 93(5):2239-2249.

PMID: 39831522 PMC: 11893242. DOI: 10.1002/mrm.30418.


Hand-held electron spin resonance scanner for subcutaneous oximetry using OxyChip.

Almog N, Zgadzai O, Kuppusamy P, Zur Y, Baruch L, Machluf M Magn Reson Med. 2024; 92(1):430-439.

PMID: 38411265 PMC: 11055674. DOI: 10.1002/mrm.30066.


EPR Monitoring of Oxygenation Levels in Tumors After Chlorophyllide-Based Photodynamic Therapy May Allow for Early Prediction of Treatment Outcome.

Szczygiel M, Kalinowska B, Szczygiel D, Krzykawska-Serda M, Fiedor L, Murzyn A Mol Imaging Biol. 2024; 26(3):411-423.

PMID: 38296885 PMC: 11211189. DOI: 10.1007/s11307-023-01886-7.


Order-of-magnitude SNR improvement for high-field EPR spectrometers via 3D printed quasi-optical sample holders.

Sojka A, Price B, Sherwin M Sci Adv. 2023; 9(38):eadi7412.

PMID: 37729398 PMC: 10511183. DOI: 10.1126/sciadv.adi7412.