The Cologne Special is a prototype of the ECAT EXACT (model 921), a new generation of Siemens-CTI PET scanners. It consists of three rings of 48 BGO block detectors each, covering an axial field of view of 16.2 cm with a patient port of 56.2 cm diameter. This results in a total of 24 rings with 384 crystals each, giving 47 contiguous image planes in two-dimensional (2D) mode. Total system sensitivity is 216 kcps/microCi/ml for a 20 cm cylinder phantom in 2D. This increases to 1.5 Mcps/microCi/ml in 3D. Data are acquired in the stationary mode only (no wobble motion), resulting in a transaxial spatial resolution of better than 6 mm full width at half-maximum (FWHM) at the center, which degrades to 7.5 mm tangentially and 9.6 mm radially at a radius of 20 cm. Average axial resolution changes from 5.0 mm FWHM at the center to 8.1 mm at R = 20 cm. Count rate performance was investigated at different low energy discriminator settings and found to be linear up to 2.5 microCi/ml with a 20 cm phantom. The magnitude and distribution of scatter were evaluated for both septa-extended and septa-retracted conditions for a range of energy thresholds. Brain, heart, and whole-body studies with the new tomograph demonstrate the versatility of its applications without compromising on physical performance.

Citing Articles

Functional Plasticity after Unilateral Vestibular Midbrain Infarction in Human Positron Emission Tomography.

Becker-Bense S, Buchholz H, Baier B, Schreckenberger M, Bartenstein P, Zwergal A PLoS One. 2016; 11(11):e0165935.

PMID: 27824897 PMC: 5100888. DOI: 10.1371/journal.pone.0165935.


Bismuth germanate coupled to near ultraviolet silicon photomultipliers for time-of-flight PET.

Kwon S, Gola A, Ferri A, Piemonte C, Cherry S Phys Med Biol. 2016; 61(18):L38-L47.

PMID: 27589153 PMC: 5056849. DOI: 10.1088/0031-9155/61/18/L38.


Reduced D2/D3 Receptor Binding of Extrastriatal and Striatal Regions in Temporal Lobe Epilepsy.

Bernedo Paredes V, Buchholz H, Gartenschlager M, Breimhorst M, Schreckenberger M, Werhahn K PLoS One. 2015; 10(11):e0141098.

PMID: 26544593 PMC: 4636381. DOI: 10.1371/journal.pone.0141098.


Quantification of dopamine transporter density with [18F]FECNT PET in healthy humans.

Nye J, Votaw J, Douglas Bremner J, Davis M, Voll R, Camp V Nucl Med Biol. 2014; 41(3):217-22.

PMID: 24533985 PMC: 4657438. DOI: 10.1016/j.nucmedbio.2013.12.013.


Ventral and dorsal streams processing visual motion perception (FDG-PET study).

Becker-Bense S, Buchholz H, Eulenburg P, Best C, Bartenstein P, Schreckenberger M BMC Neurosci. 2012; 13:81.