Safety Parameters for Pulp Temperature During Selective Ablation of Caries by KTP Laser in Vitro

Overview
Specialty General Surgery
Date 2004 May 29
PMID 15165383
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To define the optimal parameters of KTP laser irradiation during a selective caries removal.

Materials And Methods: Twelve decayed human teeth, recently extracted were used. Their root canals were prepared for insertion of a thermocouple probe into the pulp chamber. The demineralized tissues were colored by Acid Red 52 before proceeding to different conditions of irradiation.

Results: Pulpal temperature increases (below 3 degrees C) were found under the following parameters with 15 sec of continuous lasing: 400 mWatts, 0.10-msec pulse width, PRR <50 Hz for efficient caries removal. A resting time average of 70 sec was necessary to allow pulp temperature to get back to its baseline.

Conclusion: KTP laser can be used safely and without any pulp over-heating under certain irradiation conditions.

Citing Articles

Dentinal Hypersensitivity Treatment Using Diode Laser 980 nm: In Vivo Study.

El Mobadder M, Namour A, Namour M, Dib W, El Mobadder W, Maalouf E Dent J (Basel). 2019; 7(1).

PMID: 30634421 PMC: 6473910. DOI: 10.3390/dj7010005.