Stripe Wear Rates in Alumina THR--comparison of Microseparation Simulator Study with Retrieved Implants

Overview
Date 2004 Apr 30
PMID 15116404
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Alumina-on-alumina hip implants with microseparation were run in a hip simulator for comparison of the nonseparation simulator mode and retrievals. The 28-, 32-, and 36-mm Biolox-forte implants were run to 5 million cycles with the use of 50% newborn calf serum. Howmedica Osteonics Trident cups with titanium backing were used in all sets. In standard (STD) and microseparation (MSX) mode, the typical biphasic wear trend was evident, but the MSX test mode had much higher magnitudes. There was a 5-fold increase for run-in wear and up to a 35-fold increase in steady-state wear. The stripe wear on the ball formed early, but did not progress in grade beyond 0.6 Mc. The locations of the stripes were similar in retrieved and simulator balls. However, the stripes from the simulator were narrower than short-term retrievals and much narrower than some long-term retrievals. The long-term retrieved balls had a grade of wear greater than the simulators. In vivo a broader range of motion occurs and this may lead to the wider stripe observed on the retrievals. These observations suggested that simulators could produce the loading and kinematics similar to a patient walking but not necessarily the variety of motions possible in the in vivo situation.

Citing Articles

Moving fluoroscopy-based analysis of THA kinematics during unrestricted activities of daily living.

DIsidoro F, Brockmann C, Friesenbichler B, Zumbrunn T, Leunig M, Ferguson S Front Bioeng Biotechnol. 2023; 11:1095845.

PMID: 37168610 PMC: 10164959. DOI: 10.3389/fbioe.2023.1095845.


Effect of an edge at cup rim on contact stress during micro-separation in ceramic-on-ceramic hip joints.

Liu F, Fisher J Tribol Int. 2017; 113:323-329.

PMID: 28867870 PMC: 5465945. DOI: 10.1016/j.triboint.2017.01.012.


Bioceramics for Hip Joints: The Physical Chemistry Viewpoint.

Pezzotti G Materials (Basel). 2017; 7(6):4367-4410.

PMID: 28788682 PMC: 5455897. DOI: 10.3390/ma7064367.


Contact mechanics of modular metal-on-polyethylene total hip replacement under adverse edge loading conditions.

Hua X, Li J, Wang L, Jin Z, Wilcox R, Fisher J J Biomech. 2014; 47(13):3303-9.

PMID: 25218504 PMC: 4199141. DOI: 10.1016/j.jbiomech.2014.08.015.


Risk of impingement and third-body abrasion with 28-mm metal-on-metal bearings.

Clarke I, Lazennec J, Brusson A, Savisaar C, Bowsher J, Burgett M Clin Orthop Relat Res. 2013; 472(2):497-508.

PMID: 24297107 PMC: 3890183. DOI: 10.1007/s11999-013-3399-3.