» Articles » PMID: 1486811

Magnetic Resonance Imaging and Computed Tomography in the Detection and Localization of Intraocular Foreign Bodies

Overview
Journal Doc Ophthalmol
Specialty Ophthalmology
Date 1992 Jan 1
PMID 1486811
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

In this experimental study, various foreign bodies were inserted into fresh bovine eyes, in different localizations. Twenty-one magnetic and non-magnetic foreign bodies, dimensions of which varied from 1.5 x 1.5 x 2 mm to 3.5 x 6 x 7 mm, were tried to detect by computed tomography (CT) and magnetic resonance imaging (MRI) scanning. In addition, further dissections were applied to check the ocular damage attributable to movement of the foreign bodies. Ferromagnetic foreign bodies have been shown to move in the eye and the risk of torsional forces being applied to the ferromagnetic foreign body seemed to cause intraocular complications during MRI scanning. All of the foreign bodies that were implanted in bovine eyes were recognized on CT scanning, except intraocular lenses. As a general rule, metallic foreign bodies produced beam-hardening artifacts, but these artifacts did not cause any problem in detecting the localizations of foreign bodies.

Citing Articles

An Unusual Case of Low Vision and Anisocoria Considered a Neurological Finding in the Emergency Department: Ocular Siderosis.

Ozen O, Inal Ozen M Turk J Ophthalmol. 2025; 54(6):354-357.

PMID: 39743939 PMC: 11707453. DOI: 10.4274/tjo.galenos.2024.77783.


Unilateral Ocular Siderosis Bulbi Due to Missed Metallic Intraocular Foreign Body Masquerading as Anisocoria of Neurological Origin: A Case Report.

Al-Dwairi R, Msallam M Am J Case Rep. 2021; 22:e930504.

PMID: 33947824 PMC: 8112286. DOI: 10.12659/AJCR.930504.


Metallic intraocular foreign body as detected by magnetic resonance imaging without complications- A case report.

Platt A, Wajda B, Ingram A, Wei X, Ells A Am J Ophthalmol Case Rep. 2017; 7:76-79.

PMID: 29260084 PMC: 5722138. DOI: 10.1016/j.ajoc.2017.06.010.

References
1.
Weisman R, Savino P, Schut L, Schatz N . Computed tomography in penetrating wounds of the orbit with retained foreign bodies. Arch Otolaryngol. 1983; 109(4):265-8. DOI: 10.1001/archotol.1983.00800180063013. View

2.
NEW P, Rosen B, Brady T, Buonanno F, Kistler J, Burt C . Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging. Radiology. 1983; 147(1):139-48. DOI: 10.1148/radiology.147.1.6828719. View

3.
Lagouros P, Langer B, Peyman G, Mafee M, Spigos D, Grisolano J . Magnetic resonance imaging and intraocular foreign bodies. Arch Ophthalmol. 1987; 105(4):551-3. DOI: 10.1001/archopht.1987.01060040121048. View

4.
Williams S, Char D, Dillon W, Lincoff N, Moseley M . Ferrous intraocular foreign bodies and magnetic resonance imaging. Am J Ophthalmol. 1988; 105(4):398-401. DOI: 10.1016/0002-9394(88)90305-4. View

5.
Zheutlin J, Thompson J, Shofner R . The safety of magnetic resonance imaging with intraorbital metallic objects after retinal reattachment or trauma. Am J Ophthalmol. 1987; 103(6):831. DOI: 10.1016/s0002-9394(14)74404-6. View