» Articles » PMID: 14726062

Biomechanical Comparison of Median Sternotomy Closures

Overview
Journal Ann Thorac Surg
Publisher Elsevier
Date 2004 Jan 17
PMID 14726062
Citations 20
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Poor healing of median sternotomy can significantly increase morbidity, mortality, and hospital costs. Effective union requires reliable sternal fixation. Although wire has proven the most reliable and widely used sternotomy closure material, no experimental studies have compared a large variety of wiring techniques in a human model. We developed an easily reproducible experimental model using cadaveric human sterna and compared several wiring methods to assess closure strength and stability.

Methods: Fifty-three fresh adult human cadaveric sternal plates with adjacent ribs were fixed with specially designed spiked stainless steel clamps and attached to a texture analyzer. Single peristernal and transsternal, alternating single peristernal and transsternal, figure-eight peristernal, figure-eight pericostal, and Robicsek closures using no. 5 stainless steel wires were tested. We evaluated bone density, stiffness, and displacement using perpendicular, repetitive variable force loads of 800 Newtons cycling at a rate of 0.5 mm/s.

Results: There were no significant differences in age, sex, or bone density in outcome measures of the sternal groups. No clamp failures or clamp damage to the specimens occurred. The single peristernal and alternating peristernal and transsternal closures proved superior in strength and stability (p < 0.001). The figure-eight peristernal, then the single transsternal, then the Robicsek were next stablest groups in decreasing order. The figure-eight pericostal closure had the highest failure rate (p < 0.001).

Conclusions: This novel model of sternotomy closure testing was reliable, inexpensive, and easily reproducible. The mechanical stability of peristernal and alternating peristernal and transsternal wires was significantly greater than that of the other tested methods. Pericostal figure-eight closures were not sufficiently stable to be considered a reliable method of primary sternotomy repair.

Citing Articles

Improving Sternal Closure Outcomes in Cardiac Surgery: Polyethylene Suture Tapes vs. Steel Wires.

Khalpey Z, Kumar U, Aslam U, Phillips T, Khalpey Z, Cooper A J Clin Med. 2025; 14(1.

PMID: 39797360 PMC: 11720976. DOI: 10.3390/jcm14010277.


Review of Biomechanical Studies and Finite Element Modeling of Sternal Closure Using Bio-Active Adhesives.

Al-Abassi A, Papini M, Towler M Bioengineering (Basel). 2022; 9(5).

PMID: 35621476 PMC: 9138150. DOI: 10.3390/bioengineering9050198.


Clinical assessment of efficacy of poly-L-lactide sternal pin on sternal stability and post-operative pain: a prospective randomized trial in cardiovascular surgery.

Takahara S, Sasaki K, Saito T, Sakuma K, Fujiwara H, Yoshioka I J Thorac Dis. 2022; 14(1):76-89.

PMID: 35242370 PMC: 8828521. DOI: 10.21037/jtd-21-1340.


Chitosan Scaffold Containing Periostin Enhances Sternum Bone Healing and Decreases Serum Level of TNF-α and IL-6 after Sternotomy in Rat.

Salehiamin M, Toolee H, Azami M, Ahmadi Tafti S, Mojaverrostami S, Halimi S Tissue Eng Regen Med. 2022; 19(4):839-852.

PMID: 35199306 PMC: 9294132. DOI: 10.1007/s13770-022-00434-8.


Comparison of polyetheretherketone cables and stainless steel wires for sternal fixation after median sternotomy.

Shu D, Li J, Zhao Y, Yang Y J Int Med Res. 2021; 49(9):3000605211041265.

PMID: 34551599 PMC: 8485305. DOI: 10.1177/03000605211041265.