» Articles » PMID: 14646752

Surface Damage on Open Box Posterior-stabilized Polyethylene Tibial Inserts

Overview
Publisher Wolters Kluwer
Specialty Orthopedics
Date 2003 Dec 4
PMID 14646752
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Twenty retrieved Scorpio posterior-stabilized implants were available for analysis. The mean implantation time was 22 months (range, 2 days-42 months). Favorable types and amounts of surface damage were seen on the tibiofemoral and backsides of these modular PE liners that had been packaged in an inert environment and then sterilized by gamma irradiation. Delamination represented only 0.1% of the total surface damage. Off-axis loading (varus malalignment) was associated with tibial component loosening but there was no evidence of peripheral damage of PE caused by edge loading. With this open-box design, hyperextension marks on the anterior aspect of the posterior-stabilized post from femoral component impingement occurred in 11 of 20 cases and was related to sagittal component positioning: excess tibial slope or increased tibial slope combined with a flexed femoral component. Unique coarse abrasions occurred in 16 of 20 cases and were the result of cement extrusion into the open box, especially with varus malalignment. These observations provide guidance for optimizing the surgical technique and the design of posterior-stabilized total knee components.

Citing Articles

Does tibial design modification improve implant stability for total knee arthroplasty? An experimental cadaver study.

Jaeger S, Eissler M, Schwarze M, Schonhoff M, Kretzer J, Bitsch R Bone Joint Res. 2022; 11(4):229-238.

PMID: 35400170 PMC: 9057524. DOI: 10.1302/2046-3758.114.BJR-2021-0169.R1.


The John Insall Award: no functional advantage of a mobile bearing posterior stabilized TKA.

Mahoney O, Kinsey T, DErrico T, Shen J Clin Orthop Relat Res. 2011; 470(1):33-44.

PMID: 22006197 PMC: 3238000. DOI: 10.1007/s11999-011-2114-5.


A pictographic atlas for classifying damage modes on polyethylene bearings.

Harman M, Cristofolini L, Erani P, Stea S, Viceconti M J Mater Sci Mater Med. 2011; 22(5):1137-46.

PMID: 21461698 DOI: 10.1007/s10856-011-4303-x.


Damages of the tibial post in constrained total knee prostheses in the early postoperative course - a scanning electron microscopic study of polyethylene inlays.

Skwara A, Tibesku C, Reichelt R, Fuchs-Winkelmann S BMC Musculoskelet Disord. 2008; 9:83.

PMID: 18547408 PMC: 2435533. DOI: 10.1186/1471-2474-9-83.


5- to 9-year survivorship of single-radius, posterior-stabilized TKA.

Mahoney O, Kinsey T Clin Orthop Relat Res. 2008; 466(2):436-42.

PMID: 18196429 PMC: 2505140. DOI: 10.1007/s11999-007-0070-x.