Issues of Methodology, Standardization and Metabolite Recognition for 25-hydroxyvitamin D when Comparing the DiaSorin Radioimmunoassay and the Nichols Advantage Automated Chemiluminescence Protein-binding Assay in Hip Fracture Cases

Overview
Specialty Biochemistry
Date 2003 Sep 25
PMID 14503993
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Deficiency of vitamin D is commonly associated with hip fracture and treatment with vitamin D reduces hip fracture rates. Consequently, the demand for assays to measure 25-hydroxyvitamin D (25-OHD) has increased. The Nichols Advantage chemiluminescence protein-binding assay (CLPBA) for 25-OHD is a first-generation automated immunoassay with decreased turnaround time, reduced manual handling and non-radioactive label.

Methods: We compared the CLPBA to the DiaSorin radioimmunoassay (RIA) and high-performance liquid chromatography (HPLC) for the measurement of 25-OHD using 161 samples from hip fracture patients and samples before and after institution of ergocalciferol (vitamin D(2)) therapy.

Results: A negative bias for the CLPBA at concentrations below 30 nmol/L and a positive bias at 25-OHD values above 30 nmol/L compared with the RIA resulted in diagnostic discordance for one in three samples when using 30 and 50 nmol/L as decision limits. HPLC analysis confirmed the presence of a negative bias for the CLPBA at low values. Both immunoassays under-estimate 25-hydroxyvitamin D(2).

Conclusions: The discordance between 25-OHD values may be due to differences in standardization of each assay relative to HPLC. Our results emphasize the need for assay-specific clinical decision limits.

Citing Articles

Prevalence of Vitamin-D deficiency is related to severity of liver damage in Hepatitis-C patients.

Falak S, Aftab L, Saeed M, Islam A Pak J Med Sci. 2020; 36(3):445-450.

PMID: 32292450 PMC: 7150413. DOI: 10.12669/pjms.36.3.1490.


Insufficient autumn vitamin D intake and low vitamin D status in 7-year-old Icelandic children.

Bjarnadottir A, Kristjansdottir A, Hrafnkelsson H, Johannsson E, Magnusson K, Thorsdottir I Public Health Nutr. 2014; 18(2):208-17.

PMID: 24476995 PMC: 10271486. DOI: 10.1017/S1368980013003558.


Simultaneous measurement of plasma vitamin D(3) metabolites, including 4β,25-dihydroxyvitamin D(3), using liquid chromatography-tandem mass spectrometry.

Wang Z, Senn T, Kalhorn T, Zheng X, Zheng S, Davis C Anal Biochem. 2011; 418(1):126-33.

PMID: 21784054 PMC: 3164754. DOI: 10.1016/j.ab.2011.06.043.


Vitamin D in organ transplantation.

Stein E, Shane E Osteoporos Int. 2011; 22(7):2107-18.

PMID: 21207011 PMC: 4139072. DOI: 10.1007/s00198-010-1523-8.


Measurement uncertainty of 25-OH vitamin D determination with different commercially available kits: impact on the clinical cut offs.

Cavalier E, Rozet E, Gadisseur R, Carlisi A, Monge M, Chapelle J Osteoporos Int. 2009; 21(6):1047-51.

PMID: 19756833 DOI: 10.1007/s00198-009-1052-5.