A New System for Ambulatory Pulmonary Artery Pressure Recording

Overview
Journal Br Heart J
Date 1992 Aug 1
PMID 1389746
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To develop a complete system for the measurement, recording, and analysis of ambulatory pulmonary artery pressure.

Design: The new system consists of a pulmonary artery catheter, an ambulatory recorder, and a desktop computer. Pulmonary artery pressure is measured by a micromanometer tipped catheter with an in vivo calibration system to allow correction for zero drift. This catheter is plugged into a small battery powered recorder. The recorder has two input channels, one for pressure and one for an event marker. The pressure wave is sampled 32 times/s, processed by an in built computer, compressed, and stored in semi-conductor memory. On completion of a recording, data is transferred from the ambulatory recorder through a serial data link to an Acorn Archimedes desktop computer on which further data processing, statistical analysis, graphics, and printouts can be obtained.

Results: The system has been used in 18 patients, with technically successful recording in 14, less than 15 minutes of data loss in three, and 12 hours of data loss in one.

Conclusions: A new system for ambulatory pulmonary artery monitoring has been developed and used clinically with success. It may provide new perspectives on the pathophysiology of disease as it applies to everyday life.

Citing Articles

Invasive Devices and Sensors for Remote Care of Heart Failure Patients.

Radhoe S, Veenis J, Brugts J Sensors (Basel). 2021; 21(6).

PMID: 33809205 PMC: 7999467. DOI: 10.3390/s21062014.


Implantable hemodynamic monitoring devices in heart failure.

Pamboukian S, Smallfield M, Bourge R Curr Cardiol Rep. 2007; 8(3):187-90.

PMID: 17543245 DOI: 10.1007/s11886-006-0032-z.


Ambulatory pulmonary artery pressure monitoring during sleep and exercise in normal individuals and patients with COPD.

Raeside D, Brown A, Patel K, Welsh D, Peacock A Thorax. 2002; 57(12):1050-3.

PMID: 12454300 PMC: 1758791. DOI: 10.1136/thorax.57.12.1050.


Pulmonary artery pressure variation in patients with connective tissue disease: 24 hour ambulatory pulmonary artery pressure monitoring.

Raeside D, Chalmers G, Clelland J, Madhok R, Peacock A Thorax. 1999; 53(10):857-62.

PMID: 10193372 PMC: 1745077. DOI: 10.1136/thx.53.10.857.

References
1.
Levy R, Shapiro L, Wright C, Mockus L, Fox K . Syndrome X: the haemodynamic significance of ST segment depression. Br Heart J. 1986; 56(4):353-7. PMC: 1236870. DOI: 10.1136/hrt.56.4.353. View

2.
Levy R, Shapiro L, Wright C, Mockus L, Fox K . Haemodynamic response to myocardial ischaemia during unrestricted activity, exercise testing, and atrial pacing assessed by ambulatory pulmonary artery pressure monitoring. Br Heart J. 1986; 56(1):12-8. PMC: 1277380. DOI: 10.1136/hrt.56.1.12. View

3.
Levy R, Shapiro L, Wright C, Mockus L, Fox K . The haemodynamic significance of asymptomatic ST segment depression assessed by ambulatory pulmonary artery pressure monitoring. Br Heart J. 1986; 56(6):526-30. PMC: 1216399. DOI: 10.1136/hrt.56.6.526. View

4.
Levy R, Cunningham D, Shapiro L, Wright C, Mockus L, Fox K . Diurnal variation in left ventricular function: a study of patients with myocardial ischaemia, syndrome X, and of normal controls. Br Heart J. 1987; 57(2):148-53. PMC: 1277096. DOI: 10.1136/hrt.57.2.148. View

5.
Levy R, Cunningham D, Shapiro L, Wright C, Mockus L, Fox K . Continuous ambulatory pulmonary artery pressure monitoring. A new method using a transducer tipped catheter and a simple recording system. Br Heart J. 1986; 55(4):336-43. PMC: 1236735. DOI: 10.1136/hrt.55.4.336. View