Precision and Accuracy of Joint Space Width Measurements of the Medial Compartment of the Knee Using Standardized MTP Semi-flexed Radiographs

Overview
Date 2003 Sep 18
PMID 13129690
Citations 15
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To quantify the precision and accuracy of measurements of joint space width (JSW) and joint space narrowing (JSN) from the medial tibiofemoral compartment of knee radiographs using a simple and easily adaptable protocol.

Methods: Radiographs of a caliper (a surrogate for JSW) were obtained to determine the precision limits of the system under ideal conditions. Bilateral knee radiographs from 10 healthy volunteers were obtained at three different times using the metatarsophalangeal (MTP) semi-flexed view posterior-anterior position without fluoroscopy. A backlit digitizing tablet and three manual methods were used to measure JSW and analyses of precision were performed. The accuracy of measuring change in JSW (a measure of JSN) was estimated from radiographs of cadaver knees that were placed in a servo-hydraulic device that moved the femur relative to the tibia through known intervals.

Results: Radiographic measurements of the caliper inter-blade distance were comparable to the resolution limits of the backlit digitizing tablet (0.025 mm). Repeated radiography of healthy subject knees produced JSW standard deviation (SD) measurements of 0.08 mm by the median SD method, and 0.11 mm by repeated measures analysis. The accuracy of JSN measurements in the cadaver knees as a mean difference from the known reference value was 0.09 mm.

Conclusion: The results indicate a high level of precision in measurements of JSW from MTP semi-flexed view knee radiographs of normal volunteers. Reproducibility was attained through careful subject positioning without fluoroscopy and the use of a backlit digitizing tablet. From the cadaver study we can predict that greater than 0.13 mm of measured JSN represents actual or true change in JSN. This radiographic technique can be used as a primary measure for early knee osteoarthritis (OA) when cartilage thickness is decreasing and limited bony remodeling has occurred.

Citing Articles

The Role of Biomarkers in Predicting Outcomes of Anterior Cruciate Ligament Reconstruction: A Systematic Review.

Batty L, Mackenzie C, Landwehr C, Webster K, Feller J Orthop J Sports Med. 2024; 12(10):23259671241275072.

PMID: 39380669 PMC: 11460236. DOI: 10.1177/23259671241275072.


Superiority of Multiple-Joint Space Width over Minimum-Joint Space Width Approach in the Machine Learning for Radiographic Severity and Knee Osteoarthritis Progression.

Cheung J, Tam A, Chan L, Chan P, Wen C Biology (Basel). 2021; 10(11).

PMID: 34827100 PMC: 8614846. DOI: 10.3390/biology10111107.


Low Rates of Radiographic Knee Osteoarthritis 5 Years After ACL Reconstruction or Rehabilitation Alone: The Delaware-Oslo ACL Cohort Study.

Pedersen M, Grindem H, Berg B, Gunderson R, Engebretsen L, Axe M Orthop J Sports Med. 2021; 9(8):23259671211027530.

PMID: 34423060 PMC: 8375355. DOI: 10.1177/23259671211027530.


Lorecivivint, a Novel Intraarticular CDC-like Kinase 2 and Dual-Specificity Tyrosine Phosphorylation-Regulated Kinase 1A Inhibitor and Wnt Pathway Modulator for the Treatment of Knee Osteoarthritis: A Phase II Randomized Trial.

Yazici Y, McAlindon T, Gibofsky A, Lane N, Clauw D, Jones M Arthritis Rheumatol. 2020; 72(10):1694-1706.

PMID: 32432388 PMC: 7589351. DOI: 10.1002/art.41315.


Radiographic-based measurement of tibiofemoral joint space width and magnetic resonance imaging derived articular cartilage thickness are not related in subjects at risk for post traumatic arthritis of the knee.

Lonza G, Gardner-Morse M, Vacek P, Beynnon B J Orthop Res. 2019; 37(5):1052-1058.

PMID: 30908712 PMC: 6499682. DOI: 10.1002/jor.24283.