Accuracy and Repeatability of Pediatric Cardiac Output Measurement Using Doppler: 20-year Review of the Literature

Overview
Specialty Critical Care
Date 2003 Sep 5
PMID 12955181
Citations 37
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: Review of the accuracy and repeatability of Doppler cardiac output (CO) measurements in children.

Design: Publications in the scientific literature retrieved using a computerized Medline search from 1982-2002 and a manual review of article bibliographies. Studies comparing Doppler flow measurements with thermodilution, Fick, or dye dilution methods in the pediatric critical care setting were identified to assess the bias, precision, and intra- and interobserver repeatability of Doppler CO measurement. Where results were not suitable for comparison and the original measurements available, data were re-analyzed using appropriate statistical methods and presented in comparative tables.

Results: The precision of pediatric Doppler CO measurements compared to thermodilution, dye dilution, or Fick methods is around 30% and repeatability varies from less than 1% to 22%. Bias is generally less than 10% but varies considerably.

Conclusions: The bias, precision, and repeatability from study to study indicate that Doppler CO measurements are acceptably reproducible in children, with best results when used to track changes rather than absolute values, and using the transesophageal approach.

Citing Articles

The Impact of Increased PEEP on Hemodynamics, Respiratory Mechanics, and Oxygenation in Pediatric ARDS.

Junqueira F, Ferraz I, Campos F, Matsumoto T, Brandao M, Nogueira R Respir Care. 2024; 69(11):1409-1416.

PMID: 39013568 PMC: 11549636. DOI: 10.4187/respcare.12005.


Association between patent ductus arteriosus flow and home oxygen therapy in extremely preterm infants.

Termerova J, Kubena A, Liska K, Tomek V, Plavka R Pediatr Res. 2024; 96(1):208-215.

PMID: 38454005 PMC: 11257949. DOI: 10.1038/s41390-024-03120-8.


Automatic Prediction of Paediatric Cardiac Output From Echocardiograms Using Deep Learning Models.

Ufkes S, Zuercher M, Erdman L, Slorach C, Mertens L, Taylor K CJC Pediatr Congenit Heart Dis. 2023; 2(1):12-19.

PMID: 37970100 PMC: 10642111. DOI: 10.1016/j.cjcpc.2022.11.001.


Accuracy of Respiratory Variation in Inferior Vena Cava Diameter to Predict Fluid Responsiveness in Children Under Mechanical Ventilation.

Campos N, de Lima L, Ferraz I, Nogueira R, Brandao M, de Souza T Pediatr Cardiol. 2023; 45(6):1326-1333.

PMID: 36759350 DOI: 10.1007/s00246-023-03115-y.


Neonatologist Performed Echocardiography for Evaluating the Newborn Infant.

Nestaas E Front Pediatr. 2022; 10:853205.

PMID: 35402362 PMC: 8987714. DOI: 10.3389/fped.2022.853205.


References
1.
Darmon P, Hillel Z, Mogtader A, Mindich B, Thys D . Cardiac output by transesophageal echocardiography using continuous-wave Doppler across the aortic valve. Anesthesiology. 1994; 80(4):796-805; discussion 25A. DOI: 10.1097/00000542-199404000-00011. View

2.
Tibby S, Hatherill M, Marsh M, Murdoch I . Clinicians' abilities to estimate cardiac index in ventilated children and infants. Arch Dis Child. 1998; 77(6):516-8. PMC: 1717412. DOI: 10.1136/adc.77.6.516. View

3.
Sholler G, Whight C, Celermajer J . Pulsed Doppler echocardiographic assessment, including use of aortic leaflet separation, of cardiac output in children with structural heart disease. Am J Cardiol. 1986; 57(13):1195-7. DOI: 10.1016/0002-9149(86)90701-0. View

4.
Welch E, Duara S, Suguihara C, Bandstra E, Bancalari E . Validation of cardiac output measurements with noninvasive Doppler echocardiography by thermodilution and Fick methods in newborn piglets. Biol Neonate. 1994; 66(2-3):137-45. DOI: 10.1159/000244101. View

5.
Notterman D, Castello F, Steinberg C, Greenwald B, OLoughlin J, Gold J . A comparison of thermodilution and pulsed Doppler cardiac output measurement in critically ill children. J Pediatr. 1989; 115(4):554-60. DOI: 10.1016/s0022-3476(89)80280-x. View