» Articles » PMID: 12931818

Levels of Metal Ions After Small- and Large-diameter Metal-on-metal Hip Arthroplasty

Overview
Date 2003 Aug 23
PMID 12931818
Citations 38
Authors
Affiliations
Soon will be listed here.
Abstract

Metal-on-metal (MOM) bearings for hip arthroplasty are increasing in popularity. Concern remains, however, regarding the potential toxicological effects of the metal ions which these bearings release. The serum levels of cobalt and chromium in 22 patients who had undergone MOM resurfacing arthroplasty were compared with a matched group of 22 patients who had undergone 28 mm MOM total hip arthroplasty (THA). At a median of 16 months (7 to 56) after resurfacing arthroplasty, we found the median serum levels of cobalt and chromium to be 38 nmol/l (14 to 44) and 53 nmol/l (23 to 165) respectively. These were significantly greater than the levels after 28 mm MOM THA which were 22 nmol/l (15 to 87, p = 0.021) and 19 nmol/l (2 to 58, p < 0.001) respectively. Since the upper limit for normal patients without implants is typically 5 nmol/l, both groups had significantly raised levels of metal ions. MOM bearings of large diameter, however, result in a greater systemic exposure of cobalt and chromium ions than bearings of small diameter. This may be of relevance for potential long-term side-effects. It is not known to what extent this difference is due to corrosion of the surfaces of the component or of the wear particles produced.

Citing Articles

Mechanisms of Adverse Local Tissue Reactions to Hip Implants.

Eltit F, Wang Q, Wang R Front Bioeng Biotechnol. 2019; 7:176.

PMID: 31417898 PMC: 6683860. DOI: 10.3389/fbioe.2019.00176.


Particulate and ion forms of cobalt-chromium challenged preosteoblasts promote osteoclastogenesis and osteolysis in a murine model of prosthesis failure.

Yang S, Zhang K, Jiang J, James B, Yang S J Biomed Mater Res A. 2018; 107(1):187-194.

PMID: 30358096 PMC: 6366336. DOI: 10.1002/jbm.a.36553.


Biomonitoring method for the analysis of chromium and cobalt in human whole blood using inductively coupled plasma - kinetic energy discrimination - mass spectrometry (ICP-KED-MS).

Castro Georgi J, Sommer Y, Ward C, Cheng P, Jones R, Caldwell K Anal Methods. 2017; 9(23):3464-3476.

PMID: 29201158 PMC: 5709816. DOI: 10.1039/C7AY00430C.


Natural Remission of Major Periprosthetic Osteolysis following Total Hip Arthroplasty with Metal-on-Metal Bearings.

Tamaki T, Oinuma K Case Rep Orthop. 2017; 2017:2576196.

PMID: 29109884 PMC: 5646292. DOI: 10.1155/2017/2576196.


Chromium oxide coatings with the potential for eliminating the risk of chromium ion release in orthopaedic implants.

Oje A, Ogwu A R Soc Open Sci. 2017; 4(7):170218.

PMID: 28791150 PMC: 5541545. DOI: 10.1098/rsos.170218.