The Genesis Stent: A New Low-profile Stent for Use in Infants, Children, and Adults with Congenital Heart Disease

Overview
Date 2003 Jun 25
PMID 12822172
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Placement of intravascular stents that can reach adult size in infants and smaller children has been limited by the large profile and poor flexibility of currently available stents. In vitro and in vivo testing of the Genesis stent was performed to evaluate crimpability, predeployment flexibility, and radial strength. Comparisons were made to the Palmaz iliac and IntraStent (IS) LD stents. Nine physicians placed 30 Genesis stents in swine pulmonary and systemic arteries to evaluate stent deliverability/crimpability. Two swine were recovered and underwent a second catheterization 8 weeks later, where the stents (n = 8) were reexpanded to maximal size. Angiographic and intravascular ultrasound (IVUS) assessments were performed. In vitro testing revealed the Genesis stent to have superior crimpability, flexibility, and comparable radial strength to the Palmaz iliac stent, and superior crimpability and radial strength and comparable flexibility to the IS LD series. During in vivo testing, the physicians graded the Genesis stent superior to the Palmaz stent regarding crimpability and deliverability, and superior to the IS LD stent in regard to crimpability, and comparable to or superior in deliverability. In the chronic animals, the Genesis stent was expanded up to maximal diameter 8 weeks following implantation. Angiographic and IVUS revealed no fractures no in-stent restenosis. The Genesis stent can be easily delivered through smaller sheaths, which will facilitate their use in infants and smaller children with vascular stenosis.

Citing Articles

The role of the new Valeo stent in treating pulmonary artery stenoses in children with complex cardiac malformations: A report of two cases.

Sharma N, Goreczny S Am J Case Rep. 2014; 15:275-9.

PMID: 24987491 PMC: 4076226. DOI: 10.12659/AJCR.890455.


Quantification of local hemodynamic alterations caused by virtual implantation of three commercially available stents for the treatment of aortic coarctation.

Kwon S, Feinstein J, Dholakia R, LaDisa Jr J Pediatr Cardiol. 2013; 35(4):732-40.

PMID: 24259013 PMC: 3959287. DOI: 10.1007/s00246-013-0845-7.


Medium and long-term outcomes after bilateral pulmonary artery stenting in children and adults with congenital heart disease.

Gonzalez I, Kenny D, Slyder S, Hijazi Z Pediatr Cardiol. 2012; 34(1):179-84.

PMID: 22843203 DOI: 10.1007/s00246-012-0439-9.


Effectiveness and safety of percutaneous transcatheter implantation of pulmonary arterial stent in congenital heart disease.

Ko H, Kim Y, Yu J, Ko J, Park I, Seo D Korean Circ J. 2012; 42(1):40-5.

PMID: 22363382 PMC: 3283753. DOI: 10.4070/kcj.2012.42.1.40.


Ultra-high-pressure balloon angioplasty for treatment of resistant stenoses within or adjacent to previously implanted pulmonary arterial stents.

Maglione J, Bergersen L, Lock J, McElhinney D Circ Cardiovasc Interv. 2009; 2(1):52-8.

PMID: 20031693 PMC: 7199106. DOI: 10.1161/CIRCINTERVENTIONS.108.826263.