Neural Network for Photoplethysmographic Respiratory Rate Monitoring

Overview
Publisher Springer
Date 2003 Jun 14
PMID 12803287
Citations 17
Authors
Affiliations
Soon will be listed here.
Abstract

The reflection mode photoplethysmographic (PPG) signal was studied with the aim of determining respiratory rate. The PPG signal includes respiratory synchronous components, seen as frequency modulation of the heart rate (respiratory sinus arrhythmia), amplitude modulation of the cardiac pulse and respiratory-induced intensity variations (RIIVs) in the PPG baseline. PPG signals were recorded from the foreheads of 15 healthy subjects. From these signals, the systolic wave-form, diastolic waveform, respiratory sinus arrhythmia, pulse amplitude and RIIVs were extracted. Using basic algorithms, the rates of false positive and false negative detection of breaths were calculated separately for each of the five components. Furthermore, a neural network was assessed in a combined pattern recognition approach. The error rates (sum of false positive and false negative breath detections) for the basic algorithms ranged from 9.7% (pulse amplitude) to 14.5% (systolic waveform). The corresponding values for the neural network analysis were 9.5-9.6%. These results suggest the use of a combined PPG system for simultaneous monitoring of respiratory rate and arterial oxygen saturation (pulse oximetry).

Citing Articles

Remote patient monitoring strategies and wearable technology in chronic obstructive pulmonary disease.

Coutu F, Iorio O, Ross B Front Med (Lausanne). 2023; 10:1236598.

PMID: 37663662 PMC: 10470466. DOI: 10.3389/fmed.2023.1236598.


A Review of Deep Learning-Based Contactless Heart Rate Measurement Methods.

Ni A, Azarang A, Kehtarnavaz N Sensors (Basel). 2021; 21(11).

PMID: 34071736 PMC: 8198867. DOI: 10.3390/s21113719.


Sources of Inaccuracy in Photoplethysmography for Continuous Cardiovascular Monitoring.

Fine J, Branan K, Rodriguez A, Boonya-Ananta T, Ajmal , Ramella-Roman J Biosensors (Basel). 2021; 11(4).

PMID: 33923469 PMC: 8073123. DOI: 10.3390/bios11040126.


A Novel Personalized Motion and Noise Artifact (MNA) Detection Method for Smartphone Photoplethysmograph (PPG) Signals.

Tabei F, Kumar R, Phan T, McManus D, Chong J IEEE Access. 2019; 6:60498-60512.

PMID: 31263653 PMC: 6602087. DOI: 10.1109/ACCESS.2018.2875873.


Photoplethysmographic determination of the respiratory rate in acutely ill patients: validation of a new algorithm and implementation into a biomedical device.

LHer E, NGuyen Q, Pateau V, Bodenes L, Lellouche F Ann Intensive Care. 2019; 9(1):11.

PMID: 30666472 PMC: 6340913. DOI: 10.1186/s13613-019-0485-z.


References
1.
Ziege S, Schmid-Schonbein H, Grebe R, Martin E . Long-term registration of cutaneous microcirculation during general anesthesia. Int J Microcirc Clin Exp. 1998; 17(6):385-94. DOI: 10.1159/000179254. View

2.
Leon M, Lorini F . Ventilation mode recognition using artificial neural networks. Comput Biomed Res. 1998; 30(5):373-8. DOI: 10.1006/cbmr.1997.1452. View

3.
Gilbert R, AUCHINCLOSS Jr J, Brodsky J, BODEN W . Changes in tidal volume, frequency, and ventilation induced by their measurement. J Appl Physiol. 1972; 33(2):252-4. DOI: 10.1152/jappl.1972.33.2.252. View

4.
Wilks P, English M . A system for rapid identification of respiratory abnormalities using a neural network. Med Eng Phys. 1995; 17(7):551-5. DOI: 10.1016/1350-4533(95)00001-4. View

5.
Mendelson Y . Pulse oximetry: theory and applications for noninvasive monitoring. Clin Chem. 1992; 38(9):1601-7. View