Antiseptic Chamber-containing Hub Reduces Central Venous Catheter-related Infection: a Prospective, Randomized Study
Overview
Emergency Medicine
Authors
Affiliations
Objective: The hub connecting the catheter and the infusion equipment is a common portal of entry for bacteria causing catheter-related sepsis. We assessed the efficacy of a new hub model (Segur-Lock) that incorporates an antiseptic chamber filled with 3% iodinated alcohol in preventing endoluminal catheter contamination and catheter-related bloodstream infection arising at the hub.
Design: Prospective, randomized, multicenter study.
Setting: Seven medical and surgical adult intensive care units in Spain.
Patients: A total of 230 patients aged 18 yrs or older requiring the insertion of a nontunneled central venous catheter for >or=6 days from January 1, 1998, to April 30, 1999.
Interventions: Patients were randomized at the time of catheter insertion to receive catheters with standard Luer-lock connector (control group, n = 114) or catheters with the new hub model (n = 116).
Measurements And Main Results: Skin, catheter tip, and hub cultures were performed at the time the catheter was withdrawn because therapy was terminated or due to suspicion of sepsis, in which case peripheral blood and infusate cultures were simultaneously taken. Catheter-related bloodstream infection was diagnosed in 19 (8.3%) patients. Catheters were more often withdrawn because of suspicion of infection in the control group (43.8%) than in the new hub model group (30.1%, p <.035). The prevalence of culture-positive catheter hubs without associated bacteremia (colonization) was higher in the control group (14.4% vs. 4.3%, p <.001). Catheter-related bloodstream infection from the catheter hub also occurred more frequently in controls than in patients assigned to the new hub model (7% vs. 1.7%; p <.049; relative risk, 4.14; 95% confidence interval, 0.8-19).
Conclusions: This new antiseptic chamber-containing hub has proved to be effective in preventing endoluminal bacterial colonization and catheter-related bloodstream infection from hub contamination in intensive care unit patients with central venous catheters inserted for >or=6 days.
Okuda K, Yoshii Y, Yamada S, Chiba A, Hironaka I, Hori S Ann Clin Microbiol Antimicrob. 2018; 17(1):44.
PMID: 30577829 PMC: 6303997. DOI: 10.1186/s12941-018-0297-2.
The Role of Antifungals against Candida Biofilm in Catheter-Related Candidemia.
Bouza E, Guinea J, Guembe M Antibiotics (Basel). 2016; 4(1):1-17.
PMID: 27025612 PMC: 4790322. DOI: 10.3390/antibiotics4010001.
Central Venous Catheter-Related Infection in Severe Trauma Patients.
Youn S, Lee J, Kim Y, Moon J, Choi Y, Jung K World J Surg. 2015; 39(10):2400-6.
PMID: 26138873 PMC: 4554744. DOI: 10.1007/s00268-015-3137-y.
Disinfection of Needleless Connector Hubs: Clinical Evidence Systematic Review.
Moureau N, Flynn J Nurs Res Pract. 2015; 2015:796762.
PMID: 26075093 PMC: 4446481. DOI: 10.1155/2015/796762.
Influence of tracheostomy on the incidence of central venous catheter-related bacteremia.
Lorente L, Jimenez A, Martin M, Castedo J, Galvan R, Garcia C Eur J Clin Microbiol Infect Dis. 2009; 28(9):1141-5.
PMID: 19370367 DOI: 10.1007/s10096-009-0742-4.