» Articles » PMID: 12458861

A New Intraocular Lens Design to Reduce Spherical Aberration of Pseudophakic Eyes

Overview
Journal J Refract Surg
Date 2002 Dec 3
PMID 12458861
Citations 103
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The aim of this study was to design and evaluate in the laboratory a new intraocular lens (IOL) intended to provide superior ocular optical quality by reducing spherical aberration.

Methods: Corneal topography measurements were performed on 71 cataract patients using an Orbscan I. The measured corneal surface shapes were used to determine the wavefront aberration of each cornea. A model cornea was then designed to reproduce the measured average spherical aberration. This model cornea was used to design IOLs having a fixed amount of negative spherical aberration that partially compensates for the average positive spherical aberration of the cornea. Theoretical and physical eye models were used to assess the expected improvement in optical quality of an eye implanted with this lens.

Results: Measurements of optical quality provided evidence that if this modified prolate IOL was centered within 0.4 mm and tilted less than 7 degrees, it would exceed the optical performance of a conventional spherical IOL. This improvement occurred without an apparent loss in depth of focus.

Conclusion: A new IOL with a prolate anterior surface, designed to partially compensate for the average spherical aberration of the cornea, is intended to improve the ocular optical quality of pseudophakic patients.

Citing Articles

Effect of crystalline lens decentration and tilt on visual performance in eyes implanted with bifocal or extended depth of focus intraocular lenses.

Zhang Y, Zhang J, Liang C, Xu Y, Wu J, Wu J BMC Ophthalmol. 2025; 25(1):58.

PMID: 39901088 PMC: 11792569. DOI: 10.1186/s12886-025-03884-5.


Posterior Capsular Opacification and Glistening in Hydrophobic Monofocal Biaspheric Intraocular Lens Two Years After Implantation: A Case Control Study.

Hervas-Ontiveros A, Espana-Gregori E, Fresno-Canada C, Butron-Ruiz R, Cervino A J Ophthalmol. 2025; 2024():3520219.

PMID: 39781444 PMC: 11707060. DOI: 10.1155/joph/3520219.


Impact of pupil size and corneal spherical aberrations on the performance of monofocal intraocular lenses: an experimental model.

Stern B, Gatinel D Biomed Opt Express. 2024; 15(11):6433-6448.

PMID: 39553876 PMC: 11563339. DOI: 10.1364/BOE.530708.


Analysis of intraocular lens tilt and decentration after cataract surgery in eyes with high myopia using the anterior segment optical coherence tomography.

Jeon Y, Park N, Lee H, Eah K, Han J, Chung H Sci Rep. 2024; 14(1):27987.

PMID: 39543310 PMC: 11564970. DOI: 10.1038/s41598-024-78759-8.


Assessment of intraocular lens tilt and decentration after femtosecond laser-assisted and conventional cataract surgery at 12 months and beyond.

Kim B, Lee H, Jeon Y, Eah K, Park N, Chung H BMC Ophthalmol. 2024; 24(1):459.

PMID: 39428465 PMC: 11492500. DOI: 10.1186/s12886-024-03720-2.