Correction for Displacement Artifacts in 3D Phase Contrast Imaging

Overview
Date 2002 Nov 2
PMID 12412037
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To correct for displacement artifacts in 3D phase contrast imaging.

Materials And Methods: A 3D phase contrast pulse sequence was modified so that displacements of velocity measurements were restricted to one direction. By applying a postprocessing method, displaced measurements could be traced back to their accurate positions. Flow studies were performed using a phantom that generated flow through a stenosis, directed oblique relative to the phase and frequency encoding directions. Velocity profiles and streamline visualization were used to compare displaced and corrected velocity data to a reference.

Results: Velocity profiles obtained from the original measurement showed skewed profiles due to the displacement artifact, both at close proximity to the orifice as well as further downstream. After correction, concordance with the reference improved considerably.

Conclusion: The displacement artifact, which restricts the accuracy of phase contrast measurements, can be corrected for using the proposed method. Correction of the phase contrast velocity data may improve the accuracy of subsequent flow analysis and visualization.

Citing Articles

An Eulerian formulation for the computational modeling of phase-contrast MRI.

Otani T, Sekine T, Sato Y, Alves E, Wada S Magn Reson Med. 2024; 93(2):828-841.

PMID: 39270130 PMC: 11604850. DOI: 10.1002/mrm.30302.


Virtual injections using 4D flow MRI with displacement corrections and constrained probabilistic streamlines.

Roberts G, Loecher M, Spahic A, Johnson K, Turski P, Eisenmenger L Magn Reson Med. 2021; 87(5):2495-2511.

PMID: 34971458 PMC: 8884720. DOI: 10.1002/mrm.29134.


Quantitative time-of-flight MR angiography for simultaneous luminal and hemodynamic evaluation of the intracranial arteries.

Koktzoglou I, Huang R, Edelman R Magn Reson Med. 2021; 87(1):150-162.

PMID: 34374455 PMC: 8616782. DOI: 10.1002/mrm.28969.


Importance of complex blood flow in the assessment of aortic regurgitation severity using phase contrast magnetic resonance imaging.

Truedsson F, Polte C, Gao S, Johnsson A, Bech-Hanssen O, Lagerstrand K Int J Cardiovasc Imaging. 2021; 37(12):3561-3572.

PMID: 34273066 PMC: 8604828. DOI: 10.1007/s10554-021-02341-w.


Quantification of turbulence and velocity in stenotic flow using spiral three-dimensional phase-contrast MRI.

Petersson S, Dyverfeldt P, Sigfridsson A, Lantz J, Carlhall C, Ebbers T Magn Reson Med. 2015; 75(3):1249-55.

PMID: 25846511 PMC: 6618270. DOI: 10.1002/mrm.25698.