Enamel Nail Polish Does Not Interfere with Pulse Oximetry Among Normoxic Volunteers

Overview
Publisher Springer
Date 2002 Sep 6
PMID 12212998
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To determine if enamel nail polish interferes with pulse oximetry.

Design: Laboratory investigation.

Participants: 12 healthy nonsmoking volunteers.

Methods: Spectrophotometry was performed on polystyrene cuvettes painted with 3 uniform layers of nail polish enamel in triplicate. Absorbances were compared at 660 and 940 nm for 10 colors across the visible spectrum and documented by the Institute of Electrical and Electronic Engineers color code against Pantone matches on a color scanner. Colors were then selected that had the greatest A660-A940 difference and pulse oximetry (Nellcor N-209A. Pleasanton , CA) was performed on nails painted with these colors on 12 subjects using an unpainted digit as a control.

Results: When tested by the oximeter, there were no SpO2 differences detected between nail polish colors, nor were there any differences between the painted and the unpainted control digits across subjects (F = 0.51, ANOVA; p = 0.67). Blue (B), green (G), and lime green (G') possessed the largest A660-A940 difference and potentially could interfere with pulse oximetry but did not do so in the clinical model.

Conclusion: Enamel finger nail polish pigments do not interfere with pulse oximetry as previously reported, the A660-A940 difference must be greater than 1.88 +/- 0.23 SD AU in order to affect pulse oximetry.

Citing Articles

Impact of Fingernail Polish on Pulse Oximetry Measurements: A Systematic Review.

Aggarwal A, Agarwal R, Dhooria S, Prasad K, Sehgal I, Muthu V Respir Care. 2023; 68(9):1271-1280.

PMID: 37185113 PMC: 10468177. DOI: 10.4187/respcare.10399.


Accuracy and precision of pulse oximeter at different sensor locations in patients with heart failure.

Hassan A, Ahmed S, AbdelHaffeez A, Mohamed S Multidiscip Respir Med. 2021; 16(1):742.

PMID: 34316367 PMC: 8278778. DOI: 10.4081/mrm.2021.742.


The Predictive Value of Pulse Oximeters for Pulse Improvement after Angiography in Infants and Children.

Alipour M, Rastegar M, Ghaderian M, Namayandeh S, Faraji R, Pezeshkpour Z Iran J Pediatr. 2017; 26(5):e5833.

PMID: 28203338 PMC: 5297377. DOI: 10.5812/ijp.5833.


Pulse oximetry in bronchiolitis: is it needed?.

Hendaus M, Jomha F, Alhammadi A Ther Clin Risk Manag. 2015; 11:1573-8.

PMID: 26491341 PMC: 4608625. DOI: 10.2147/TCRM.S93176.


Effects of passive smoking on students at College of Applied Medical Sciences, King Saud Bin Abdulaziz University for Health Sciences, Riyadh.

Alanazi A, Enezi F, Alqahtani M, Alshammari T, Ansari M, Al-Oraibi S J Nat Sci Biol Med. 2015; 6(1):100-5.

PMID: 25810644 PMC: 4367017. DOI: 10.4103/0976-9668.149100.


References
1.
Peters S . The effect of acrylic nails on the measurement of oxygen saturation as determined by pulse oximetry. AANA J. 1997; 65(4):361-3. View

2.
New Jr W . Pulse oximetry. J Clin Monit. 1985; 1(2):126-9. DOI: 10.1007/BF02832200. View

3.
Eisenkraft J . Pulse oximeter desaturation due to methemoglobinemia. Anesthesiology. 1988; 68(2):279-82. DOI: 10.1097/00000542-198802000-00018. View

4.
Kataria B, Lampkins R . Nail polish does not affect pulse oximeter saturation. Anesth Analg. 1986; 65(7):824. DOI: 10.1213/00000539-198607000-00029. View

5.
White P, Boyle W . Nail polish and oximetry. Anesth Analg. 1989; 68(4):546-7. View