Calibration of New High Dose Rate 192Ir Sources

Overview
Journal Med Phys
Specialty Biophysics
Date 2002 Aug 1
PMID 12148729
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

The increased popularity of high dose rate (HDR) brachytherapy in the treatment of neoplastic disease has resulted in the introduction of new or redesigned 192Ir sources by various manufacturers. In order to provide accurate clinical dosimetry, accurate calibration of these sources is required. In this work, two new sources are calibrated with the standard, NIST traceable, in-air calibration technique. These sources are the Varian VariSource (VS2000) and the newly redesigned Nucletron MicroSelectron source. An uncertainty budget was performed to determine the overall uncertainty of the measurement, and a total relative expanded uncertainty (k = 2) of 2.15% was calculated. The results of the source calibration using the in-air technique and that obtained based upon the 1991 standard calibration all lie well within this uncertainty. In addition to the in-air calibration, contact autoradiographs of the sources were made with International Specialty Products model HD810 radiochromic film to demonstrate the sources relative dose distributions.

Citing Articles

Dosimetric evaluation of newly developed well-type ionization chamber for use in the calibration of brachytherapy sources.

Saminathan S, Godson H, Ponmalar R, Manickam R, Mazarello J J Med Phys. 2017; 41(4):234-239.

PMID: 28144115 PMC: 5228046. DOI: 10.4103/0971-6203.195187.


Ir-192 Calibration in Air with Farmer Chamber for HDR Brachytherapy.

Chang L, Ho S, Lee T, Ding H, Chen P J Med Biol Eng. 2016; 36:145-152.

PMID: 27231462 PMC: 4853450. DOI: 10.1007/s40846-016-0117-0.


A feasibility study of Fricke dosimetry as an absorbed dose to water standard for 192Ir HDR sources.

deAlmeida C, Ochoa R, Lima M, David M, Pires E, Peixoto J PLoS One. 2014; 9(12):e115155.

PMID: 25521914 PMC: 4270754. DOI: 10.1371/journal.pone.0115155.


Calibration of (192)Ir high dose rate brachytherapy source using different calibration procedures.

Bondel S, Ravikumar M, Supe S, Reddy B Rep Pract Oncol Radiother. 2014; 19(3):151-6.

PMID: 24944818 PMC: 4059024. DOI: 10.1016/j.rpor.2013.07.014.


Estimation of distance error by fuzzy set theory required for strength determination of HDR (192)Ir brachytherapy sources.

Kumar S, Datta D, Sharma S, Chourasiya G, Babu D, Sharma D J Med Phys. 2014; 39(2):85-92.

PMID: 24872605 PMC: 4035620. DOI: 10.4103/0971-6203.131281.