» Articles » PMID: 11792836

High Throughput DNA Sequencing with a Microfabricated 96-lane Capillary Array Electrophoresis Bioprocessor

Overview
Specialty Science
Date 2002 Jan 17
PMID 11792836
Citations 31
Authors
Affiliations
Soon will be listed here.
Abstract

High throughput DNA sequencing has been performed by using a microfabricated 96-channel radial capillary array electrophoresis (microCAE) microchannel plate detected by a 4-color rotary confocal fluorescence scanner. The microchannel plate features a novel injector for uniform sieving matrix loading as well as high resolution, tapered turns that provide an effective separation length of 15.9 cm on a compact 150-mm diameter wafer. Expanded common buffer chambers for the cathode, anode, and waste reservoirs are used to simplify electrode addressing and to counteract buffering capacity depletion arising from the high electrophoretic current. DNA sequencing data from 95 successful lanes out of 96 lanes run in parallel were batch-processed with basefinder, producing an average read length of 430 bp (phred q > or = 20). Phred quality values were found to exceed 40 (0.01% probability of incorrectly calling a base) for over 80% of the read length. The microCAE system demonstrated here produces sequencing data at a rate of 1.7 kbp/min, a 5-fold increase over current commercial capillary array electrophoresis technology. Additionally, this system permits lower reagent volumes and lower sample concentrations, and it presents numerous possibilities for integrated sample preparation and handling. The unique capabilities of microCAE technology should make it the next generation, high performance DNA sequencing platform.

Citing Articles

Nanofluidic devices for the separation of biomolecules.

Rathnayaka C, Amarasekara C, Akabirov K, Murphy M, Park S, Witek M J Chromatogr A. 2022; 1683:463539.

PMID: 36223665 PMC: 9795076. DOI: 10.1016/j.chroma.2022.463539.


Diagnostic Value of Next-Generation Sequencing in Periprosthetic Joint Infection: A Systematic Review.

Tang Y, Zhao D, Wang S, Yi Q, Xia Y, Geng B Orthop Surg. 2021; 14(2):190-198.

PMID: 34935279 PMC: 8867422. DOI: 10.1111/os.13191.


Next-Generation Sequencing and the CRISPR-Cas Nexus: A Molecular Plant Virology Perspective.

Shahid M, Sattar M, Iqbal Z, Raza A, Al-Sadi A Front Microbiol. 2021; 11:609376.

PMID: 33584572 PMC: 7874184. DOI: 10.3389/fmicb.2020.609376.


Micro-flow analysis by molecular tagging velocimetry and planar Raman-scattering.

Roetmann K, Schmunk W, Garbe C, Beushausen V Exp Fluids. 2020; 44(3):419-430.

PMID: 32214636 PMC: 7087862. DOI: 10.1007/s00348-007-0420-1.


A new DNA-based model for finite field arithmetic.

Jiron I, Soto S, Marin S, Acosta M, Soto I Heliyon. 2020; 5(12):e02901.

PMID: 31890936 PMC: 6926258. DOI: 10.1016/j.heliyon.2019.e02901.


References
1.
Woolley A, Mathies R . Ultra-high-speed DNA sequencing using capillary electrophoresis chips. Anal Chem. 1995; 67(20):3676-80. DOI: 10.1021/ac00116a010. View

2.
Giddings M, Severin J, Westphall M, Wu J, Smith L . A software system for data analysis in automated DNA sequencing. Genome Res. 1998; 8(6):644-65. PMC: 310740. DOI: 10.1101/gr.8.6.644. View

3.
Shi Y, Simpson P, Scherer J, Wexler D, Skibola C, Smith M . Radial capillary array electrophoresis microplate and scanner for high-performance nucleic acid analysis. Anal Chem. 1999; 71(23):5354-61. DOI: 10.1021/ac990518p. View

4.
Kheterpal I, Scherer J, Clark S, Radhakrishnan A, Ju J, Ginther C . DNA sequencing using a four-color confocal fluorescence capillary array scanner. Electrophoresis. 1996; 17(12):1852-9. DOI: 10.1002/elps.1150171209. View

5.
Paegel B, Hutt L, Simpson P, Mathies R . Turn geometry for minimizing band broadening in microfabricated capillary electrophoresis channels. Anal Chem. 2000; 72(14):3030-7. DOI: 10.1021/ac000054r. View