Long-term Vessel Response to a Self-expanding Coronary Stent: a Serial Volumetric Intravascular Ultrasound Analysis from the ASSURE Trial.A Stent Vs. Stent Ultrasound Remodeling Evaluation

Overview
Date 2001 Apr 13
PMID 11300443
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: We sought to investigate the in vivo mechanical properties of a new self-expanding coronary stent (RADIUS) and, particularly, the subsequent vessel response over time.

Background: Preclinical studies have suggested that self-expanding stents may produce less vessel wall injury at initial deployment, leading to larger follow-up lumens than with balloon-expandable stents. However, the influence of the chronic stimulus from self-expanding stents on the vessel wall remains unknown.

Methods: Sixty-two patients were randomly assigned to either the RADIUS self-expanding stent group (n = 32) or the Palmaz-Schatz balloon-expandable stent group (n = 30). Intravascular ultrasound was performed after stent deployment and at six-month follow-up.

Results: At follow-up, the RADIUS stents had increased 23.6% in overall volume, while the Palmaz-Schatz stents had remained unchanged. Due to the greater mean neointimal area (3.0 +/- 1.7 mm2 vs. 1.9 +/- 1.2 mm2, p = 0.02) in the RADIUS group, no significant difference in net late lumen loss was observed between the two groups. On the other hand, analysis at the peristent margins demonstrated that mean late loss was significantly smaller in the RADIUS group than it was in the Palmaz-Schatz group (0.1 +/- 2.1 mm2 vs. 1.9 +/- 2.4 mm2, p = 0.02).

Conclusions: Serial volumetric IVUS revealed that the RADIUS stents continued to enlarge during the follow-up period. In this stent implantation protocol, this expansion was accompanied by a greater amount of neointima than the Palmaz-Schatz stents, resulting in similar late lumen loss in both configurations. In the peristent margins, however, late lumen loss was minimized with the RADIUS stents.

Citing Articles

Heterogeneous vascular response after implantation of bare nitinol self-expanding stents in the swine femoropopliteal artery.

Kawarada O, Otsuka F, Miki K, Ikutomi M, Okada K, Ogata S Cardiovasc Interv Ther. 2022; 38(2):210-222.

PMID: 36255689 PMC: 10020252. DOI: 10.1007/s12928-022-00889-5.


Long-term changes after carotid stenting assessed by intravascular ultrasound and near-infrared spectroscopy.

Stechovsky C, Hajek P, Roland R, Horvath M, Veselka J Cardiovasc Diagn Ther. 2022; 11(6):1180-1189.

PMID: 35070788 PMC: 8748489. DOI: 10.21037/cdt-21-160.


Outcome of Wingspan Stent Using Aggressive Post-stent Balloon Dilation for Intracranial Atherosclerosis Stenosis.

Perng P, Sun Y, Wang H, Shih Y, Lee J, Wang L Front Neurol. 2021; 12:757175.

PMID: 34759885 PMC: 8572975. DOI: 10.3389/fneur.2021.757175.


Reduction in incomplete stent apposition area caused by jailed struts after single stenting at left main bifurcation lesions: micro-CT analysis using a three-dimensional elastic bifurcated coronary artery model.

Hikichi Y, Umezu M, Node K, Iwasaki K Cardiovasc Interv Ther. 2016; 32(1):12-17.

PMID: 26968169 PMC: 5214711. DOI: 10.1007/s12928-016-0380-6.


Causes, assessment, and treatment of stent thrombosis--intravascular imaging insights.

Ong D, Jang I Nat Rev Cardiol. 2015; 12(6):325-36.

PMID: 25781415 DOI: 10.1038/nrcardio.2015.32.